Temperature display

OPERATING INSTRUCTIONS

Temperature display

The unit is set at the factory to obtain the proper temperature automatically (4oC±1oC). There is no need to set the temperature.

Table 1. Temperature display

: Lamp ON : Lamp OFF

 

 

Operation

Key operated

Display

 

Display mode

 

Plug-in the power

 

 

 

 

Average monitor bottle

UPPER

 

1

supply cord and turn on

-----

 

 

Average temperature

 

 

temperature

LOWER

 

the power switch.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Press the temperature

 

 

 

 

Upper monitor bottle

UPPER

 

2

 

UPPER

Upper temperature

display key.

 

 

 

 

temperature

LOWER

 

LOWER

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Press the temperature

 

 

 

Lower monitor bottle

UPPER

 

3

 

UPPER

Lower temperature

display key.

 

 

 

 

temperature

LOWER

 

 

LOWER

 

 

 

 

 

 

 

 

 

 

 

 

Press the temperature

 

 

 

Average monitor bottle

UPPER

 

4

 

UPPER

Average temperature

display key.

 

 

 

 

temperature

LOWER

 

 

LOWER

 

 

 

 

 

 

 

 

 

 

 

5

Repeat from step 2.

 

 

 

 

 

 

 

Note:

The displayed temperature represents the temperature sensed by a thermal sensor attached to the monitor bottle mounted to the upper and lower portion of the chamber.

The displayed temperature may not always be at 4oC. Because the chamber probes are representative of high and low average temperatures within the chamber.

Defrost cycles

There is no need for routine defrosting of the unit as this occurs automatically as follows:

1. Cycle defrost

To keep the chamber temperature stable, the refrigeration compressor is cycled on and off. During offperiod any frost which has accumulated on the evaporator is melted by energizing a defrost heater. This will not have any discernible effect on the chamber temperature.

2. Forced defrost

When the ambient humidity is high, or a large amount of damp product is being stored inside the chamber, there is a possibility that cycle defrost may not be enough to remove all of the frost on the evaporator. In this case, a forced defrost cycle can be initiated.

When the unit is operating under a forced defrost cycle, the current chamber temperature and dFis displayed alternately on the digital temperature display.

Once the forced defrost cycle is completed, normal operation resumes.

The temperature of blood bag (400mL) in the chamber rises by approximately 1 to 2oC during the defrosting.