ALARMS & SAFETY FUNCTIONS

This unit has the alarms and safety functions shown in Table 2, and also self diagnostic functions.

Table 2 Alarms and safety functions

Alarm & Safety

Situation

Indication

Buzzer

Safety operation

High temperature

If the upper or lower temperature sensor

Alarm lamp flashes.

Intermittent tone

Remote alarm

alarm

senses a temperature 6oC.

Low temperature

If the upper or lower temperature sensor

Alarm lamp flashes.

Intermittent tone

Remote alarm

alarm

senses a temperature 2oC.

Protection of

If the high-temperature thermostat

 

 

Compressor ON.

excessive high

sensor senses a temperature

-----

-----

Heater OFF.

temperature

approximately > 10oC.

 

 

Protection of

If the alarm sensor senses a

 

 

Compressor ON.

excessive low

-----

-----

temperature approximately < 2oC.

Heater OFF.

temperature

 

 

 

 

 

When the power to the unit is

 

 

 

Power failure alarm

disconnected or power switch is

Alarm lamp flashes.

Intermittent tone

Remote alarm.

 

turned off.

 

 

 

 

 

 

Intermittent tone

 

Door alarm

When the door is ajar or open.

Door check lamp is lit.

with 2 minutes

-----

 

 

 

delay.

 

Auto-return

When there is no key pressing in

Monitor bottle temperature is

-----

Finishing of calibration

each setting mode for 90 seconds.

displayed.

mode.

 

 

 

If the temperature sensor of upper

Alarm lamp flashes.

 

 

 

monitor bottle is disconnected or

E1 and monitor bottle temp.

Intermittent tone

Remote alarm.

 

short-circuited.

are displayed alternately.

 

 

 

If the temperature sensor of lower

Alarm lamp flashes.

 

 

 

monitor bottle is disconnected or

E2 and monitor bottle temp.

Intermittent tone

Remote alarm.

Thermal sensor

short-circuited.

are displayed alternately.

 

 

abnormality

If the defrost sensor is disconnected or

Alarm lamp flashes.

 

 

 

E3 and monitor bottle temp.

Intermittent tone

Remote alarm.

 

short-circuited.

 

are displayed alternately.

 

 

 

 

 

 

 

If the over-temperature sensor is

Alarm lamp flashes.

 

 

 

E4 and monitor bottle temp.

Intermittent tone

Remote alarm.

 

disconnected or short-circuited.

 

are displayed alternately.

 

 

 

 

 

 

Note:

The remote alarm is silenced by pressing the buzzer stop key (BUZZER) as the remote alarm is operated in conjunction with the buzzer, except for the power failure alarm. All alarm functions will auto-reset. The buzzer and remote contacts will re-activate if the condition which caused the alarm is not corrected within 30 minutes.

Alarm conditions will remain active for about 24 hours on a fully charged back-up battery in the event of a power failure. Two days operation is required to fully charge the battery after initial start-up if the power has been off for an extended period.

The remote alarm contacts will remain active even if the battery is discharged.

Alarm test procedure

The alarm function can be checked by the following procedure:

1.Press alarm test (ALARM/TEST) key.

2.The audible and visual alarm indicators activate along with the remote alarm. The alarm condition remains active for 30 seconds and then terminates automatically. The audible alarm (buzzer) cannot be silenced by pressing the buzzer stop key (BUZZER).

3.Otherwise the alarm test can be terminated manually at any time during the rest by pressing the ALARM/TEST key again.