SPECIFICATIONS
Name |
|
| Blood Bank Refrigerator |
| ||
Model |
|
| ||||
External dimensions |
|
| W600 x D680 x H1835 (mm) |
| ||
Internal dimensions |
|
| W520 x D490 x H1150 (mm) |
| ||
Effective capacity |
| 304 L | 302 L |
| ||
Exterior |
|
| Acrylic finish baked on zinc galvanized steel |
| ||
Interior |
|
| Acrylic finish baked on zinc galvanized steel |
| ||
Door |
|
| 1 door with glass window |
| ||
Insulation |
|
| Rigid polyurethane |
| ||
Shelf | Hard steel wire on polyethylene coating |
| ||||
|
| 5 pcs. |
|
|
| |
Drawer |
| Stainless steel, Handle with card holder | ||||
|
|
|
| 5 pcs. |
| |
Access port |
|
| Left side, Inner diameter 30 mm |
| ||
Cooling method |
|
| Forced cool air circulation |
| ||
Compressor |
|
| Hermetic type, Output; 175 W |
| ||
Condenser |
|
| Fin and tube type |
| ||
Evaporator |
|
| Fin and tube type |
| ||
Refrigerant |
|
|
| |||
Temperature controller |
|
| Electronic control system |
| ||
Temperature display |
|
| Digital display |
| ||
Alarm | High temperature, Low temperature, Door, Power failure, Sensor | |||||
Remote alarm contact |
|
| Allowable contact capacity: 30 VDC, 2 A |
| ||
Memory |
|
| Nonvolatile static memory |
| ||
Fluorescent lamp |
|
| White light 15 W x 1 (Glow starter; |
| ||
Temperature recorder |
|
| ||||
Accessories |
|
| 1 set of key |
| ||
Battery |
|
| 6 VDC, 1100 mAh |
| ||
Weight |
| 125 kg | 140 kg |
|
Note: The contents of this manual are subject to change without notice.
PERFORMANCE
Monitor bottle temperature |
|
|
|
| +4oC±1oC |
|
|
|
|
Usable ambient temp. |
|
|
|
| +5 to 35oC |
|
|
|
|
Power failure alarm |
|
| 24 hours (when the battery is full charged) |
|
| ||||
duration |
|
|
|
|
|
|
|
|
|
Power source | 110V | 115V |
| 220V | 220 V | 230 V |
| 240 V | |
| 60Hz | 60Hz |
| 50Hz | 60Hz | 50Hz |
| 50Hz | 50Hz |
Rated power consumption | 220W | 220W |
| 220W | 220W | 220W |
| 220W | 220W |
Noise level |
|
|
| 41 dB [A] (background noise; 20 dB) |
|
| |||
Maximum pressure |
|
|
|
| 1275 kPa |
|
|
|
|
Note :The unit with CE mark complies with EC directives 89/336/EEC, 93/68/EEC and 73/23/EEC.
23