Instruction Manual

8.SPECIFICATIONS

Model

4105-02

4105-03

4105-04

4105-05

4105-06

4105-07

4105-10

Gas

Comb.

O2

CO

H2S

Cl2

H2

SO2

Sensor Type

Catalytic

E.C.

E.C.

E.C

E.C.

E.C.

E.C.

Units

%LEL

% Vol

ppm

ppm

ppm

ppm

ppm

Range

0-100

0-25%

0-1000

0-100

0-10

0-1000

0-100

Resolution

3% F.S.

0.1% F.S.

1

0.1

0.1

2

0.5

Response Time

Note 1

10 sec.

25 sec.

30 sec.

60 sec.

30 sec.

20 sec.

to 90% of signal

 

 

 

 

 

 

 

Sensor Life 2

3 Yrs

2 yrs

2 yrs

2 yrs

2 yrs

2 yrs

2 yrs

Operating Range

 

 

 

 

 

 

 

Temperature

14 to 158oF

19 to 122oF

14 to 122oF

14 to 122oF

14 to 122oF

14 to 122oF

14 to 122oF

 

-10 to 75oC

-7 to 50oC

-10 to 50oC

-10 to 50oC

-10 to 50oC

-10 to 50oC

-10 to 50oC

Relative Humidity

10-95%

5-99%

15-90%

15-90%

15-90%

10-90%

15-90%

Pressure

10%

10%

10%

10%

10%

10%

10%

Electrical Data

 

 

 

 

 

 

 

Input Voltage DC

 

 

19-30VDC at less than 100 mA

 

 

Output-Normal

 

 

4-20 mA DC linear

 

 

Output-Trouble

 

 

 

0 mA

 

 

 

Construction

 

 

 

 

 

 

 

Dimensions

 

H: 7.6", D: 4.6", W: 4.2" (19.3 x 11.7 x 10.7 cm)

 

Weight (Module)

 

 

 

5.0 lb. (2.2 kg)

 

 

Mounting

 

 

 

3/4" NPT

 

 

 

Housing

 

 

Explosion proof

 

 

 

 

(NEMA 7) / NEMA 4X Enclosure (Div. I, Class 1, Groups B, C, D)

 

Notes: 1. Step to 50% LEL within 10 sec, recovery to 10% LEL within 30 sec.

2.Sensor life is for use at standard temperature and pressure with occasional exposure to the gas of interest

Page: 10

Model 4105 Gas Sensor Module (12/02)

Page 13
Image 13
Sierra Monitor Corporation 4105 manual Specifications