Instruction Manual

8.SPECIFICATIONS (Cont.) Cross Sensitivities (toxic sensors)

Model

Gas

Reading from 100 ppm of interfering gas

 

 

 

 

 

Number

Type

CO

H2S Cl2

H2

SO2

NO2

NO

HCl

HCN

C2H4

4105-04

CO

100

315

-15

<40

50

-55

30

2

40

<50

4105-05

H2S

<0.5

100

-20

<0.1

<15

-15

0

0

0

0

4105-06

Cl2

0

<-10

100

0

0

105

0

0

0

0

4105-07

H2

<1

<20

0

100

3

0

35

3

35

85

4105-10

SO2

<1

0

-40

0

100

-100

0

0

15

0

4105-12

N O 2

0

-20

90

0

0

100

0

0

-3

0

4105-21

HCl

0

75

-10

0

35

-2

0

100

-8

0

4105-22

HCN

<0.5

 

-50

0

160

-190

-5

30

100

<1

4105-25

N H

0

0

0

-

40

0

0

-

-

-

 

3

 

 

 

 

 

 

 

 

 

 

4105-26

HF

0

0

9

0

9

5

-

6

-

-

Page: 12

Model 4105 Gas Sensor Module (12/02)

Page 15
Image 15