8.SPECIFICATIONS (Cont.) Cross Sensitivities (toxic sensors)
Model | Gas | Reading from 100 ppm of interfering gas |
|
|
|
|
| ||||
Number | Type | CO | H2S Cl2 | H2 | SO2 | NO2 | NO | HCl | HCN | C2H4 | |
CO | 100 | 315 | <40 | 50 | 30 | 2 | 40 | <50 | |||
H2S | <0.5 | 100 | <0.1 | <15 | 0 | 0 | 0 | 0 | |||
Cl2 | 0 | 100 | 0 | 0 | 105 | 0 | 0 | 0 | 0 | ||
H2 | <1 | <20 | 0 | 100 | 3 | 0 | 35 | 3 | 35 | 85 | |
SO2 | <1 | 0 | 0 | 100 | 0 | 0 | 15 | 0 | |||
N O 2 | 0 | 90 | 0 | 0 | 100 | 0 | 0 | 0 | |||
HCl | 0 | 75 | 0 | 35 | 0 | 100 | 0 | ||||
HCN | <0.5 |
| 0 | 160 | 30 | 100 | <1 | ||||
N H | 0 | 0 | 0 | - | 40 | 0 | 0 | - | - | - | |
| 3 |
|
|
|
|
|
|
|
|
|
|
HF | 0 | 0 | 9 | 0 | 9 | 5 | - | 6 | - | - |
Page: 12 | Model 4105 Gas Sensor Module (12/02) |