5. OPERATION

Model 4501-XX Toxic Gas Sensor Module

5. OPERATION

5.1INTRODUCTION

The 4501-XX Toxic Gas Sensor Module utilizes a visual menu system operated by means of a magnet. A magnet stick is supplied for this purpose. The menu system is used to calibrate the sensor module, and for maintenance and diagnostic procedures.

The following adjustments can be made to the sensor module:

Range adjustments (see section 5.4)

4mA and 20mA calibration adjustments (see section 5.4)

Span adjustment (see section 6.3.1)

5.2DATA ENTRY KEYPAD

The module menu system is operated by means of directing the magnet stick toward each of four independent

hall-effect magnetic switches. Each switch functions as if it is a manually activated panel key. The keys are located under the faceplate above and below the LCD and are labeled M , E , S and T as shown in Figure 5-1.

Key •Key  ENTER: MODE

Key Figure 5-1 ENTER

Key Manual background: UP (+)

Key Manual background: DOWN (-)

Figure 5-1

Face Plate with Operator Keypad