Sierra Monitor Corporation 4501-05 Model 4501-XXToxic Gas Sensor Module, Sub-RoutineA, Function

Models: 4501-05 2-Wire Series 4501-XX 4501-04 4501-06

1 25
Download 25 pages 29.36 Kb
Page 14
Image 14
Model 4501-XX Toxic Gas Sensor Module

 

 

 

 

 

Model 4501-XX Toxic Gas Sensor Module

 

5.4 MAINTENANCE SUB-MENU

 

 

 

 

 

 

 

 

 

 

 

 

Key

Function

Display

Description

Reference

 

 

 

 

 

H2S

Default Display (example 4501-05)

 

 

 

 

 

 

XX PPM

 

 

 

 

 

 

 

 

 

 

M

E

Mode

H2S

 

 

 

 

S

T

CALIB:

 

 

 

 

 

 

 

 

 

 

 

 

M

E

Down

H2S

 

 

 

 

S

T

MAINT:

 

 

 

 

 

 

 

 

 

 

 

 

M

E

Enter

Cal: 4mA

Enter Calibration 4 mA Menu

 

 

 

S

T

Banner: Use [S] or [T] to change 4mA output, [E] when done

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

A multi-meter should be inserted into the current loop

 

 

 

 

 

 

 

to measure the loop current between the power and

 

 

 

 

 

 

 

the sensor module No gas is required for this step

 

 

 

M

E

Enter

4mA

Acknowledges 4mA calibration

or Sub-Routine A

 

 

S

T

DONE

 

 

 

 

 

 

 

 

 

 

M

E

Down

H2S

 

 

 

 

S

T

MAINT:

 

 

 

 

 

 

 

 

 

 

 

 

M

E

Enter

MAINT

Enter Calibration 20 mA Menu

 

 

 

 

 

CAL:20mA

Banner: Use [S] or [T] to change 20mA output, [E] when

 

 

 

S

T

 

 

 

 

done

 

 

 

 

 

 

 

 

 

 

 

 

A multi-meter should be inserted into the current loop

 

 

 

 

 

 

 

to measure the loop current between the power and

 

 

 

 

 

 

 

the sensor module No gas is required for this step

 

 

 

M

E

Enter

20mA

Acknowledges 20mA calibration

or Sub-Routine A

 

 

S

T

DONE

 

 

 

 

 

 

 

 

 

 

 

 

 

H2S

 

 

 

 

 

 

 

MAINT:

 

 

 

 

 

 

 

 

 

 

 

M

E

Down

MAINT:

 

 

 

 

S

T

4mA

 

 

 

 

 

 

 

 

 

 

 

 

M

E

Down

MAINT:

 

 

 

 

S

T

20mA

 

 

 

 

 

 

 

 

 

 

 

 

M

E

Enter

H2S

Enter Range Menu

 

 

 

S

T

Range

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

M- Range

Enter Calibration Range Menu

 

 

 

 

 

 

100 PPM

Banner: Use [S] or [T] to change range, [E] when done

 

 

 

 

 

 

 

 

 

 

 

M

E

Enter

Range

Acknowledge range selection

 

 

 

S

T

DONE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Sub-Routine A

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CAL: XmA

The expected range of the 4mA and 20mA is limited to factory

 

 

 

 

 

 

defined values. If the actual signal is outside of this

 

 

 

 

 

 

ERROR!

predefined range, an error is displayed indicating a hardware

 

 

 

 

 

 

 

or installation problem. Contact technical support.

 

 

 

 

 

 

 

Table 5-3

 

 

 

 

 

 

 

Maintenance Sub Menu

 

 

 

 

 

 

 

 

 

 

Page: 10

Page 14
Image 14
Sierra Monitor Corporation 4501-05 Model 4501-XXToxic Gas Sensor Module, Sub-RoutineA, Function, Display, Description