Sierra Monitor Corporation 2-Wire Series Model 4501-XXToxic Gas Sensor Module, Page, Key Function

Models: 4501-05 2-Wire Series 4501-XX 4501-04 4501-06

1 25
Download 25 pages 29.36 Kb
Page 18
Image 18
6.3 DIAGNOSTICS SUB-MENU

 

 

 

 

 

 

Model 4501-XX Toxic Gas Sensor Module

 

6.3 DIAGNOSTICS SUB-MENU

 

 

 

 

 

 

 

 

 

 

 

Key Function

Display
Description
Reference

 

M

E

Mode

H2S

 

Default Display (example 4501-05)

or Sub-Routine A

 

S

T

CALIB:

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

M

E

Enter

CAL 0 PPM

 

Banner: Apply zero gas, [E] when done

 

 

S

T

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Operation: Confirm area clear of gas, or apply zero air to sensor.

 

 

 

 

 

 

 

 

 

 

M

E

Enter

Cal0PPM

 

There is a minimum time to calibrate at 0PPM. If [E] is hit prior to this time

 

 

S

T

Wait.XXX

 

has elapsed, a countdown screen will appear.

 

 

 

 

 

ZERO

 

Zero gas setting completed (Message will appear on the screen and after 3

 

 

 

 

 

DONE

 

seconds will go to the next menu - 25PPM Span)

 

 

 

 

 

XXPPM

 

Banner: [S] or [T] to change span value, [E] when done

Sub-Routine B

 

 

 

 

 

 

 

 

 

 

 

 

H2S

 

Operation: 148 second time out before sensor is returned to service.

 

 

 

 

 

WAIT-148

 

 

 

 

 

 

 

 

 

 

M

E

Enter

H2S

 

Banner: Remove gas from sensor

Sub-Routine C

 

S

T

CAL-OK

 

 

 

 

 

 

 

Sub-Routine A - Abort Calibration

 

 

 

 

 

 

 

 

 

 

M

E

Mode

CALIB

 

Operation: Hold magnet over [M] for 10 seconds to abort calibration

 

 

S

T

ABORTED

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

H2S

 

Default Display

 

 

 

 

 

XXX PPM

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Sub-Routine B - Select Span

 

 

 

 

 

 

 

 

 

 

 

25PPM SPAN

Banner: [S] or [T] to change span value, [E] when done

 

 

 

 

 

 

 

 

 

M

E

Down

C##PPM

 

Banner: Apply xx-PPM gas, [E] when stable

Should be stable after 3

 

S

T

 

minutes of gas exposure

 

M

E

Enter

H2S

 

Operation: Remove gas from sensor

If calibration failed, then

 

S

T

REMOVE GAS

 

Sub-Routine C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

H2S

 

Calibration successful

 

 

 

 

 

CAL-OK

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Sub-Routine C - Calibration Fail

 

 

 

 

 

 

 

 

 

 

M

E

Enter

READ

 

Operation: Remove gas from sensor read manual, particularly section 6.3 and

 

 

S

T

MANUAL

 

7.6, to check calibration procedures.

 

 

 

 

 

H2S

 

Calibration Failed

 

 

 

 

 

CAL-FAIL

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Table 6-1

 

 

 

 

 

 

 

Calibration Sub-Menu

 

Page: 14

Page 18
Image 18
Sierra Monitor Corporation 2-Wire Series Model 4501-XXToxic Gas Sensor Module, Page, Diagnostics Sub-Menu, Key Function