2 | QUIRC User Guide |
in cases where the background has been changing rapidly. It may also give better results if the dome flat was not evenly illuminated (it is difficult to achieve even illumination at the 0.6m telescope). The dark should be subtracted from the sky before division. The disadvantages of this technique are that the sky flat shows the response of the detector to the OH airglow + thermal emission. In particular, fringing may be present in certain configurations (e.g. 2.2m f/10 1:1), and fringing is something which should be subtracted, not divided.
The number of bad pixels usually dictates a special technique for observing in which several exposures are made of the field being studied, with each exposure shifted slightly from the others (dithering). When the images are combined, the bad pixels in one image can be “filled in" with good pixels from a shifted image. This technique also improves
It is recommended that at the start and end of each night dome flats and darks be taken. The darks should ideally be exposures of the same length as the object exposures. Even if the darks are not directly used in the reduction, they will serve to show which pixels have high dark counts so that these pixels can be included in a bad pixel mask. Dome flats are generally taken as a lights on/lights off pair. Using this strategy results in a difference image (ON – OFF) which represents the detector’s
The shutter is a leaf type shutter, meaning than the center part of the aperture is open slightly longer than the outside. Recent tests showed significant
At the 2.2m telescope there is a slight rotation in the nominal cassegrain rotator position (270). The rotation was measured in February 1996 to be 0.883 degrees CCW (e.g., N is rotated 0.883 deg E of vertical when displayed in the normal way). One could attempt to adjust slightly for this by changing the rotator position, or adjust for it later during data reduction. If the precise rotation and scale is important to the observations, one must measure this carefully during the run since the exact rotation value is likely to change slightly between runs when the instrument is taken off the telescope and remounted.
2.1Detector Linearity, Saturation, Read Noise, Dark Current
Hard saturation of the detector occurs at 50,000 ADU’s. The total gain of the system results in a scale factor of 1.85 electrons/ADU. Recent tests (2/96) showed the device to be linear to better than 1% for values up to about 44,000 ADUs. However, the gain and illumination is variable across the array so care must be taken so that parts of the array are not saturating when the average ADU value is getting close to the