User calibration

With fixed reference points

0 point / 1/3 load / Max. load

Switch the scale off. Switch the scale on.

Press the and keys and hold for 2 sec. during power-on.

Empty the scale and confirm the zero point by pressing .

Place the default weight on the scale (1/3 load) and con- firm by pressing ..

Place the default weight on the scale (Max. load) and confirm by pressing .

Empty the scale.

The scale goes to weighing mode.

Nullstellen

With variable reference point 0 point Test weight < Max

Switch the scale off. Switch the scale on.

Press the and keys and hold for 2 sec. when switching on.

Empty the scale and confirm the zero point by pressing

the key.

You can toggle between the decades by pressing the

and keys.

You can increment the numerical value by pressing the key.

Enter the existing weight, e.g. 6kg, and confirm by pres- sing .

Place the existing weight on the scale and confirm by pressing .

Empty the scale.

The scale goes to weighing mode.