Owner's Record
The model and serial numbers are located at the rear. Record these numbers in the space provided below. Refer to these numbers whenever you call upon your Sony dealer regarding this product.
Model No. ____________________
Serial No. ____________________
WARNING
To prevent fire or shock hazard, do not expose the unit to rain or moisture.
To avoid electrical shock, do not open the cabinet. Refer servicing to qualified personnel only.
THIS APPARATUS MUST BE EARTHED.
This symbol indicates the equipotential terminal which brings the various parts of a system to the same potential.
This symbol is intended to alert the user to the presence of important operating and maintenance (servicing) instructions in the literature accompanying the appliance.
Important safeguards/notices for use in the medical environments
1.All the equipments connected to this unit shall be certified according to Standard
2.When this unit is used together with other equipment in the patient area*, the equipment shall be either powered by an isolation transformer or connected via an additional protective earth terminal to system ground unless it is certified according to Standard
*Patient Area
3.The leakage current could increase when connected to other equipment.
4.This equipment generates, uses, and can radiate frequency energy. If it is not installed and used in accordance with the instruction manual, it may cause interference to other equipment. If this unit causes interference (which can be determined by
unplugging the power cord from the unit), try these measures: Relocate the unit with respect to the susceptible equipment. Plug this unit and the susceptible equipment into different branch circuit. Consult your dealer. (According to standard
Caution
When you dispose of the unit or accessories, you must obey the law in the relative area or country and the regulation in the relative hospital.
Warning on power connection
Use a proper power cord for your local power supply.
Warning on power connection for medical use
Please use the following power supply cord.
With connectors (plug or female) and cord types other than those indicated in this table, use the power supply cord that is approved for use in your area.
| United States | Canada |
|
|
|
Plug Type | HOSPITAL GRADE* | HOSPITAL GRADE* |
|
|
|
Female end | E62405, E35708 | LR53182, LL022442, |
|
| LL088408 |
Cord type | E159216, E35496 | |
| Min.Type SJT | LL32121, LL84494 |
| Min.18 AWG | Min.Type SJT |
|
| Min.18AWG |
|
|
|
Minimum cord set | 10A/125V | 10A/125V |
rating |
|
|
Safety approval | UL Listed | CSA |
|
|
|
*Note: Grounding reliability can only be achieved when the equipment is connected to an equivalent receptacle marked ‘Hospital Only’ or ‘Hospital Grade’.
2