Guidance and manufacturer’s declaration - electromagnetic immunity
The
Immunity test | IEC 60601 test | Compliance level | Electromagnetic | |
level | ||||
|
|
| ||
|
|
|
| |
Electrostatic | ±6 kV contact | ±6 kV contact | Floors should be wood, concrete or ceramic tile. If floors are | |
discharge (ESD) |
|
| covered with synthetic material, the relative humidity should | |
IEC | ±8 kV air | ±8 kV air | be at least 30%. | |
|
|
| ||
|
|
|
| |
Electrical fast | ±2 kV for power | ±2 kV for power | Mains power quality should be that of a typical commercial or | |
transient/burst | supply lines | supply lines | hospital environment. | |
IEC | ±1 kV for input/ | ±1 kV for input/ |
| |
| output lines | output lines |
| |
|
|
|
| |
Surge | ±1 kV differential | ±1 kV differential | Mains power quality should be that of a typical commercial or | |
IEC | mode | mode | hospital environment. | |
±2 kV common | ±2 kV common |
| ||
|
| |||
| mode | mode |
| |
|
|
|
| |
Voltage dips, short | < 5% UT | < 5% UT | Mains power quality should be that of a typical commercial or | |
interruptions and | (> 95% dip in UT) | (> 95% dip in UT) | hospital environment. If the user of the | |
voltage variations | for 0.5 cycle | for 0.5 cycle | continued operation during power mains interruptions, it is | |
on power supply | recommended that the | |||
|
| |||
input lines | 40% UT | 40% UT | uninterruptible power supply or a battery. | |
|
| |||
IEC | (60% dip in UT) | (60% dip in UT) |
| |
| for 5 cycles | for 5 cycles |
| |
| 70% UT | 70% UT |
| |
| (30% dip in UT) | (30% dip in UT) |
| |
| for 25 cycles | for 25 cycles |
| |
| < 5% UT | < 5% UT |
| |
| (> 95% dip in UT) | (> 95% dip in UT) |
| |
| for 5 sec | for 5 sec |
| |
|
|
|
| |
Power frequency | 3 A/m | 3 A/m | Power frequency magnetic fields should be at least | |
(50/60 Hz) |
|
| characteristic of a typical location in a typical commercial or | |
magnetic field |
|
| hospital environment. |
IEC
NOTE: UT is the a.c. mains voltage prior to application of the test level.
4