Guidance and manufacturer’s declaration - electromagnetic immunity

The UP-DR80MD is intended for use in the electromagnetic environment specified below. The customer or the user of the UP-DR80MD should assure that it is used in such as environment.

Immunity test

IEC 60601 test

Compliance level

Electromagnetic environment-guidance

level

 

 

 

 

 

 

 

Electrostatic

±6 kV contact

±6 kV contact

Floors should be wood, concrete or ceramic tile. If floors are

discharge (ESD)

 

 

covered with synthetic material, the relative humidity should

IEC 61000-4-2

±8 kV air

±8 kV air

be at least 30%.

 

 

 

 

 

 

 

Electrical fast

±2 kV for power

±2 kV for power

Mains power quality should be that of a typical commercial or

transient/burst

supply lines

supply lines

hospital environment.

IEC 61000-4-4

±1 kV for input/

±1 kV for input/

 

 

output lines

output lines

 

 

 

 

 

Surge

±1 kV differential

±1 kV differential

Mains power quality should be that of a typical commercial or

IEC 61000-4-5

mode

mode

hospital environment.

±2 kV common

±2 kV common

 

 

 

 

mode

mode

 

 

 

 

 

Voltage dips, short

< 5% UT

< 5% UT

Mains power quality should be that of a typical commercial or

interruptions and

(> 95% dip in UT)

(> 95% dip in UT)

hospital environment. If the user of the UP-DR80MD requires

voltage variations

for 0.5 cycle

for 0.5 cycle

continued operation during power mains interruptions, it is

on power supply

recommended that the UP-DR80MD be powered from an

 

 

input lines

40% UT

40% UT

uninterruptible power supply or a battery.

 

 

IEC 61000-4-11

(60% dip in UT)

(60% dip in UT)

 

 

for 5 cycles

for 5 cycles

 

 

70% UT

70% UT

 

 

(30% dip in UT)

(30% dip in UT)

 

 

for 25 cycles

for 25 cycles

 

 

< 5% UT

< 5% UT

 

 

(> 95% dip in UT)

(> 95% dip in UT)

 

 

for 5 sec

for 5 sec

 

 

 

 

 

Power frequency

3 A/m

3 A/m

Power frequency magnetic fields should be at least

(50/60 Hz)

 

 

characteristic of a typical location in a typical commercial or

magnetic field

 

 

hospital environment.

IEC 61000-4-8

NOTE: UT is the a.c. mains voltage prior to application of the test level.