Recommended separation distances between portable and mobile RF communications equipment and the
The
Rated maximum output power of transmitter | Separation distance according to frequency of transmitter | ||
W |
| m |
|
|
|
|
|
| 150 kHz to 80 MHz | 80 MHz to 800 MHz | 800 MHz to 2.5 GHz |
| d = 1.2 √P | d = 1.2 √P | d = 2.3 √P |
|
|
|
|
0.01 | 0.12 | 0.12 | 0.23 |
|
|
|
|
0.1 | 0.38 | 0.38 | 0.73 |
|
|
|
|
1 | 1.2 | 1.2 | 2.3 |
|
|
|
|
10 | 3.8 | 3.8 | 7.3 |
|
|
|
|
100 | 12 | 12 | 23 |
|
|
|
|
For transmitters rated a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Caution
When you dispose of the unit or accessories, you must obey the law in the relative area or country and the regulation in the relative hospital.
Warning on power connection
Use a proper power cord for your local power supply.
1.Use the approved Power Cord
2.Use the Power Cord
If you have questions on the use of the above Power Cord
/Appliance Connector / Plug, please consult a qualified service personnel.
Warning on power connection for medical use
Please use the following power supply cord.
With connectors (plug or female) and cord types other than those indicated in this table, use the power supply cord that is approved for use in your area.
| United States and Canada |
|
|
Plug Type | HOSPITAL GRADE* |
|
|
Cord Type | Min.Type SJT |
| Min.18 AWG |
Minimum Rating for Plug | 10A/125V |
and Appliance Couplers |
|
|
|
Safety Approval | UL Listed and CSA |
|
|
*Note: Grounding reliability can only be achieved when the equipment is connected to an equivalent receptacle marked ‘Hospital Only’ or ‘Hospi- tal Grade’.
For the customers in Europe
The manufacturer of this product is Sony Corporation, 1-
The Authorized Representative for EMC, medical devices, and product safety is Sony Deutschland GmbH, Hedelfinger Strasse 61, 70327 Stuttgart, Germany.
For any service or guarantee matters, please refer to the addresses given in separate service or guarantee documents.
6