set the value by 0,1 hPa or 0,01 inHg steps. The reference value should be measured with at least the same accuracy. If you are uncertain, contact the Suunto technical service before changing the calibration.

To change the pressure sensor calibration:

1.In the Function menu, scroll to General and press Suunto. The General menu is displayed.

2.Scroll to Sensor and press Suunto. The text Sensor Calibration appears and the current measured absolute air pressure is displayed as active.

3.With the arrow buttons, set the absolute air pressure value to match the measured reference value and press Suunto. You return to the Function menu.

NOTE: You can return the original calibration by scrolling the sensor calibration value until text DEFAULT is displayed in the middle row.

3.2.6. Units

The units menu includes the units that are used with all the Suunto X6/X6M functions and settings. The unit settings affect the way information is displayed in all the Suunto X6/X6M modes.

Setting the time format

The time setting defines whether the 12- or 24-hour clock is used.

To change the 12/24-hour display:

1.In the Function menu, scroll to Units and press Suunto. The Units menu is displayed.

2.Scroll to Time and press Suunto. The 12/24 h field is activated.

3.Select the correct setting with the arrow buttons and press Suunto. The time display is set.

22

Page 24
Image 24
Suunto X6m manual Units, Setting the time format