8.Troubleshooting
WProblem List
If problems occur, please refer to the following instructions before you ask for repairs.
Problem | Solution |
RS-232C Interface Instructions
This section describes the interface for outputting the | GB |
mCAUTION Only connect
Nothing appears on the display after | • Please confirm that the Control Box is properly connected to the AC | ||
pushing [ | ] key |
| Adapter and the AC Adapter is properly connected to the cable. |
|
| ||
| • | ||
|
|
| compared to height and weight. |
|
|
| Please make sure that measurement was carried out with bare feet, the |
|
|
| electrodes or the soles of the feet were clean, and the feet were correctly in |
|
|
| contact with the electrodes. If the feet are dry or have calluses, place water on |
|
|
| the electrodes before stepping on the unit. |
|
| • | It is possible that the body fat ratio exceeds the measuring capacity. |
|
|
| Please stop measuring. |
|
|
| |
• | This sign is displayed when the body fat percentage of the trunk is too | ||
the trunk cannot be obtained. |
| low and is off the measureable range. |
W Specifications
Communication Standard | EIA |
|
|
Communication Method | Asynchronous Transmission |
|
|
Signal speed | 4800 baud |
|
|
Data bit length | 8 bits |
|
|
Parity | None |
|
|
Stop bit | 1 bit |
|
|
Terminator | CR+LF |
|
|
WSignal lines and connection methods
Terminal number | Signal name |
|
|
1 | F.G. |
2 | RXD |
3 | TXD |
5 | GND |
1 23 5
| • | Please refer to the readings of other parts for the body fat. |
|
|
|
• | E - 11 is displayed when there is a loose connection between the control | |
|
| box and the weighing platform. |
| • Confirm that none of connections between the scale and control box are | |
|
| loose or unplugged. |
| • There may be excessive vibration which will disturb the measurement process. | |
|
| |
• The equipment requires readjustment. | ||
|
|
|
• | Measurement is not possible because impedance is unstable. In order to | |
|
| improve electrical conductivity, please clean the soles of the feet or |
|
| place water on the electrodes before using the equipment. |
|
| |
No print out | • Please confirm the number of print outs selected is more than 0. (See P. 13) | |
| • | Please confirm that the correct brand of paper is used. |
| • | Please confirm that the printer paper roll is placed in the correct position. |
| • | Please confirm that the printer paper is not jammed. |
|
|
|
The printer paper does not come out. | • | Please confirm the printer paper is not jammed. |
|
| |
P.End is displayed and the equipment | • The printer paper is used up. Please put in a new printer paper roll. (See P.11) | |
is incapable of taking a measurement. | • | Please confirm that the printer paper is being advanced properly. |
| • Please make sure that the Paper Release Lever is not in the “Up” position. (See P. 25) | |
|
| |
• It is possible that the maximum weight capacity has been exceeded. | ||
|
| Please stop measuring. |
|
| |
uuuu is displayed. | • Do not stand on or put things on the Weighting Platform before starting measurement. | |
The power has turned off during measurement. |
| Start measurement after confirming there is nothing on the Weighing Platform. |
|
|
|
Feed Key is not functioning. | • | It is possible that the number of print outs selected is 0 or the Weight |
|
| Scale Mode has been selected. Please select the Body Fat Measurement |
|
| Mode and select a number larger than 0 for the number of print outs. |