TW-357 REV E
Lid Switch
Is attached to the hinge and determines whether or not the lid is open on the freezer. This also allows the control to determine whether to activate the Quickchill, Auto Defog or Lid Alarm features.
Lid Switch
Figure 3.0 Lid Switch
Solenoid Valve
Designed to work with 12 VDC solenoid valves manufactured by Valcor,
Thermocouples
Type T thermocouples determine the temperature in the freezer. The user may choose to use NONE, 1 or 2 thermocouples with this control at any time. (The unit comes complete with one Thermocouple)
Wall Transformer
A 16.5 VAC, 40 VA wall transformer is a standard option on the LABS cryostorage system with battery backup. A 24 VAC, 40 VA wall transformer is supplied for the Kryos version (No Battery). The system is supplied with a transformer compatible with common household (North American) 110VAC. (For other power outlets contact
Battery Backup
The LABS cryostorage system with the battery backup option has a 12VDC battery installed underneath the unit's tabletop.
Remote Alarm
If an error condition occurs after a user defined period of time a remote alarm circuit can be initiated. This is accomplished by connecting a remote device to the remote alarm jack on the rear electrical panel. The
Figure 4.0 Remote Alarm Plug Connection
13