Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1

Table 2.5 - Enter the Expected Span Gas Concentrations Procedure

Step No.

Action

Comment

 

 

 

 

1.

Press

This key sequence causes the M201A to

prompt for the expected TNX concentration.

 

 

 

CAL-

Enter the TNX span concentration value by

 

CONC-

pressing the key under each digit until the

 

TNX

expected value is set. This menu can also be

 

 

entered from CALS or CALZ.

2.

Press

ENTR stores the expected TNX span value.

 

ENTR

 

3.

Press

Now enter the expected TN span concentration

 

CAL-

as in step one.

 

 

 

CONC-TN

 

 

 

 

4.

Press

Pressing ENTR stores the TN span value and

 

ENTR

returns the prompt to the CONC menu.

5.

Press

Returns instrument to SAMPLE mode.

 

EXIT

 

 

 

 

 

 

 