Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1
Table 2.5 - Enter the Expected Span Gas Concentrations Procedure
Step No. | Action | Comment | |
| |||
|
|
| |
1. | Press | This key sequence causes the M201A to | |
prompt for the expected TNX concentration. | |||
|
| ||
| CAL- | Enter the TNX span concentration value by | |
| CONC- | pressing the key under each digit until the | |
| TNX | expected value is set. This menu can also be | |
|
| entered from CALS or CALZ. | |
2. | Press | ENTR stores the expected TNX span value. | |
| ENTR |
| |
3. | Press | Now enter the expected TN span concentration | |
| CAL- | as in step one. | |
|
| ||
|
| ||
|
|
| |
4. | Press | Pressing ENTR stores the TN span value and | |
| ENTR | returns the prompt to the CONC menu. | |
5. | Press | Returns instrument to SAMPLE mode. | |
| EXIT |
| |
|
|
| |
|
|
|
14