Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1

Table 2.3 - Span Calibration Procedure - Span Gas thru Sample Port

Step No.

Action

Comment

 

1.

Press CAL

The M201A enters the calibrate mode. Input NO span

 

 

gas through the sample port.

2.

Press

Set the converter efficiency parameter to 1.0000, then

 

CAL-CONV-SET

press ENTR-EXIT.

 

 

 

2.

Wait 20 min

Wait for reading to stabilize at span value.

 

 

 

3.

Press SPAN

If you change your mind after pressing SPAN, you can

 

 

still press EXIT here without spanning the instrument.

4.

Press ENTR

Pressing ENTR actually changes the calculation

 

 

equations and causes the instrument to read the TN

 

 

and TNx span concentrations.

5.

Press EXIT

M201A returns to sampling. Immediately after

 

 

calibration, data is not added to the DAS averages.

 

 

 