Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1
Table 2.3 - Span Calibration Procedure - Span Gas thru Sample Port
Step No. | Action | Comment |
| ||
1. | Press CAL | The M201A enters the calibrate mode. Input NO span |
|
| gas through the sample port. |
2. | Press | Set the converter efficiency parameter to 1.0000, then |
| press | |
|
|
|
2. | Wait 20 min | Wait for reading to stabilize at span value. |
|
|
|
3. | Press SPAN | If you change your mind after pressing SPAN, you can |
|
| still press EXIT here without spanning the instrument. |
4. | Press ENTR | Pressing ENTR actually changes the calculation |
|
| equations and causes the instrument to read the TN |
|
| and TNx span concentrations. |
5. | Press EXIT | M201A returns to sampling. Immediately after |
|
| calibration, data is not added to the DAS averages. |
|
|
|
12