![Part No.](/images/new-backgrounds/49362/4936239x1.webp)
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1
Table 3.2 – M501NH Converter Rebuild Parts List
Part No.
KIT000191
KIT000139
HE0000007
Description
Converter Rebuild Kit
Ceramic Bushings
Ceramic Heater 220W 60 VAC
Procedure:
1.Turn off power to the converter and allow to cool. IT IS VERY IMPORTANT THAT
THE CONVERTER BE COOL BEFORE ATTEMPTING ANY DISASSEMBLY OR REPAIRS.
CAUTION VERY HOT
Will cause severe burns – disassembly while hot will damage other converter components. Allow sufficient time to cool.
2.Remove the chassis cover of the M501NH, remove the aluminum cover over the oven, remove the two
3.Disconnect the 1 stainless steel and 2 PTFE tubes from the converter cartridge.
4.Loosen the nut holding the cartridge assembly in the
5.Gently lift and tilt the converter assembly out of the
6.Loosen the ¼” SS tube fitting nearest the
7.
19