Teledyne 201A manual 2.1.2 TN/TNx Channel Calibration Quality, Step No, Action, Comment

Models: 201A

1 31
Download 31 pages 62.78 Kb
Page 16
Image 16
2.1.2 TN/TNx Channel Calibration Quality

Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1

Table 2.6 - CALS - Span Cal Procedure - Gas from rear panel span gas inlet

Step No.

Action

Comment

 

1.

Press

The M201A enters the calibrate mode. The

 

CALS

Z/S valves are activated to route span gas into

 

 

the analyzer.

2.

Press

Set the converter efficiency parameter to 1.0000, then

 

CAL-CONV-SET

press ENTR-EXIT.

 

 

 

2.

Wait 20

Wait for the reading to stabilize at span value.

 

min

 

3.

Press

If you change your mind after pressing SPAN,

 

SPAN

you can still press EXIT here without spanning

 

 

the instrument.

4.

Press

Pressing ENTR actually changes the

 

ENTR

calculation equations and causes the

 

 

instrument to read the TN and TNx span

 

 

concentrations.

5.

Press EXIT

M201A returns to sampling. Immediately after

 

 

calibration, data is not added to the DAS

 

 

averages.

 

 

 

2.1.2 TN/TNx Channel Calibration Quality

After the TN and TNx channel calibration it is important to check the slopes and offsets to make sure they fall within the acceptable range. The tolerances are different than a typical NOx analyzer primarily because of the performance of the M501NH ammonia converter.

Slopes can vary from 2.0 to 0.7, and should be within 20% of each other with the TNx slope typically smaller than the TN slope.

Offsets should be zero ±250 mV and be within 100mV of each other.

If the slope and offsets are within specifications, proceed on the ammonia channel calibration.

15