![1.2 M201A Analyzer Specifications](/images/new-backgrounds/49362/4936219x1.webp)
Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1
1.2 M201A Analyzer Specifications
Ranges | |
Range Modes | Single, Independent, AutoRange |
Noise at Zero1 | 0.5 ppb RMS |
Noise at Span1 | < 1.0% of reading above 50 ppb |
Lower Detectable Limit | 1 ppb RMS |
Zero Drift3 | 2 ppb / 24 hours |
Span Drift | < 1.0% FS Range / 7 days |
Lag Time | 40 sec |
Rise Time | 90% 120 sec |
Fall Time | 90% 120 sec |
Sample Flow Rate | 1000 cc/min |
Linearity | 2% of full scale |
Precision | 1.0% of reading |
Temp Range | |
Dimensions HxWxD | 7” x 17” x 23.6” (18cm x 43cm x 61cm) |
Weight, Analyzer | 43 lbs (20 kg) |
Weight, Converter | 24 lbs (11 kg) |
Weight, Pump | 16 lbs (7 kg) |
Power, Analyzer | 100V ~50/60 Hz, 120V ~60 Hz, 220V ~50Hz, 240V ~50 Hz, |
| 125 watts |
Power, Analyzer5 | 230V ~50 Hz, 125 watts |
Power, Pump | 110V ~60 Hz, 220V ~50 Hz, 240V ~50 Hz, 295 watts |
Power, Pump CEMark5 | 230 V ~50 Hz, 2.5 A peak |
Environmental | Installation Category Pollution Degree 2, |
| Category II |
Analog Output | |
Analog Resolution | 1 part in 2048 |
Status Option | 12 Status Outputs |
Measurement Units | ppb, ug/m3 |
1.As defined by USEPA
2.Defined as twice the zero noise level
3.At constant temperature and voltage
4.
5.Electrical rating for CE Mark Compliance
9