Teledyne manual 1.2 M201A Analyzer Specifications

Models: 201A

1 31
Download 31 pages 62.78 Kb
Page 10
Image 10
1.2 M201A Analyzer Specifications

Teledyne API Model 201A NH3 Analyzer Instruction Manual, 04689, Rev. B1

1.2 M201A Analyzer Specifications

Ranges

0-50 to 0-2000 ppb in 1 ppb increments

Range Modes

Single, Independent, AutoRange

Noise at Zero1

0.5 ppb RMS

Noise at Span1

< 1.0% of reading above 50 ppb

Lower Detectable Limit

1 ppb RMS

Zero Drift3

2 ppb / 24 hours

Span Drift

< 1.0% FS Range / 7 days

Lag Time

40 sec

Rise Time

90% 120 sec

Fall Time

90% 120 sec

Sample Flow Rate

1000 cc/min

Linearity

2% of full scale

Precision

1.0% of reading

Temp Range

10-30oC

Dimensions HxWxD

7” x 17” x 23.6” (18cm x 43cm x 61cm)

Weight, Analyzer

43 lbs (20 kg)

Weight, Converter

24 lbs (11 kg)

Weight, Pump

16 lbs (7 kg)

Power, Analyzer

100V ~50/60 Hz, 120V ~60 Hz, 220V ~50Hz, 240V ~50 Hz,

 

125 watts

Power, Analyzer5

230V ~50 Hz, 125 watts

Power, Pump

110V ~60 Hz, 220V ~50 Hz, 240V ~50 Hz, 295 watts

Power, Pump CEMark5

230 V ~50 Hz, 2.5 A peak

Environmental

Installation Category Pollution Degree 2, Over-voltage

 

Category II

Analog Output

0-100, 0-1, 0-5, 0-10VDC bi-polar, 4-20mA isolated

Analog Resolution

1 part in 2048

Status Option

12 Status Outputs

Measurement Units

ppb, ug/m3

1.As defined by USEPA

2.Defined as twice the zero noise level

3.At constant temperature and voltage

4.Bi-polar

5.Electrical rating for CE Mark Compliance

9

Page 10
Image 10
Teledyne manual 1.2 M201A Analyzer Specifications