MODEL GFC7000EU CO ANALYZER WITH
(Addendum to GFC7000E Manual PN 04584)
| Humidity Range | |
| Temp Coefficient | < 0.05 % of reading per °C (5 ppb/°C minimum)(6) |
| Voltage Coefficient | < 0.05 % of reading per V |
| Dimensions | 7" x 17" x 23.5" (178 mm x 432 mm x 597 mm) |
| (HxWxD) |
|
| Weight | 50 lb (22.7 kg) |
| AC Power | 100V 50/60 Hz (3.25A), 115 V 60 Hz (3.0A), |
|
| 220 – 240 V 50/60 Hz (2.5A) |
| Environmental | Installation Category (Over voltage Category) II |
| Conditions | Pollution Degree 2 |
| Analog Outputs | Three (3) Outputs |
| Analog Output | 100 mV, 1 V, 5 V, 10 V, |
| Ranges | current loop. |
|
| All Ranges with 5% Under/Over Range |
| Analog Output | 1 part in 4096 of selected |
| Resolution |
|
| Status Outputs | 8 Status outputs from |
| Control Inputs | 6 Control Inputs, 2 defined, 4 spare |
| I/O | One (1) |
|
| Baud Rate : 300 - 115200 |
| Certifications | CE: |
|
| A |
(1)As defined by the USEPA
(2)At constant temperature and voltage
(3)Or 0.2 ppm, whichever is greater
(4)Or 0.1 ppm, whichever is greater
(5)Above 10 ppm range, otherwise 0.2 ppm for lower ranges
3.1.2. Pneumatic Connections
This section is similar to the one in the manual, with the following exception. The extra port labeled “From Purge” is the exhaust port for the for the purge gas. The gas needs to exhausted with a gas line not more than 10 meters long.
In addition, the GFC7000EU uses a different sample pressure sensor than the GFC7000E. The GFc7000EU pressure sensor is rated for oxygen service. The part number is listed in the appendix – spare parts.
6.13.1.3Relay alarm outputs
3 | 05134 Rev - |