MODEL GFC7000EU CO ANALYZER WITH
(Addendum to GFC7000E Manual PN 04584)
| Warning | Fault Condition | Possible Causes |
| Message | ||
|
|
| |
|
|
|
|
| FRONT PANEL | The CPU is unable to | WARNING only appears on Serial I/O COM Port(s) |
| WARN | Communicate with | Front Panel Display will be frozen, blank or will not |
|
| the Front Panel | respond. |
|
| Display /Keyboard | Failed Keyboard |
|
|
| |
|
|
| I2C Bus failure |
|
|
| Loose Connector/Wiring |
| PHOTO TEMP | PHT DRIVE is | Failed IR |
| WARNING | >4750 mVDC or | Failed Sync/Demod Board |
|
| <250 MVDC | IR |
|
| Sample Chamber | |
|
|
| |
|
|
| BENCH TEMP too high. |
| REAR BOARD | Mother Board not | WARNING only appears on Serial I/O COM Port(s) |
| NOT DET | detected on power | Front Panel Display will be frozen, blank or will not |
|
| up. | respond. |
|
|
| Massive Failure of Mother Board |
| RELAY BOARD | The CPU cannot | I2C Bus failure |
| WARN | communicate with | Failed Relay Board |
|
| the Relay Board. | Loose connectors/wiring |
|
|
|
|
| PURGE PRESS | Purge Pressure | Loss of purge flow |
| WARN | outside of warning | Failed purge pressure sensor |
|
| limits specified by | Purge pressure regulator not set properly |
|
| PURGE_PRESS_SET |
|
|
| variable |
|
|
|
|
|
| SAMPLE PRES | Sample Pressure is | If Sample Pressure is < 15 |
| WARN | <15 | Blocked Particulate Filter |
|
| > 35 | Blocked Sample Inlet/Gas Line |
|
| Normally 29.92 in- | Failed Pressure Sensor/circuitry |
|
| Hg at sea level | If Sample Pressure is > 35 |
|
| decreasing at 1 in- | Blocked Vent line on pressurized |
|
| Hg per 1000 ft of | |
|
| Sample/Zero/Span gas supply | |
|
| altitude (with no | |
|
| Bad Pressure Sensor/circuitry | |
|
| flow – pump | |
|
| Failed Permapure dryer | |
|
| disconnected). | |
|
|
| |
|
|
|
|
| SAMPLE TEMP | Sample temperature | Ambient Temperature outside of specified |
| WARN | is < 10oC or > | range |
|
| 100oC. | Failed Bench Heater |
|
|
| Failed Bench Temperature Sensor |
|
|
| Relay controlling the Bench Heater |
|
|
| Failed Relay Board |
|
|
| I2C Bus |
| SOURCE | Occurs when CO Ref | GFC Wheel stopped |
| WARNING | is <1250 mVDC or | Failed Sync/Demod Board |
|
| >4950 mVDC. | If Status LED’s on the Sync/Demod Board ARE |
|
|
| flashing the cause is most likely a failed: |
|
| Either of these | IR Source |
|
| conditions will result | Relay Board |
|
| in an invalid M/R | I2C Bus |
|
| ratio. | IR |
|
|
|
|
| SYNC | Phase Lock Loop | GFC Wheel stopped |
| WARNING | (PLL) has lost lock | GFC Wheel rotation very slow/dragging |
|
|
|
|