MODEL GFC7000EU CO ANALYZER WITH AUTO-REFERENCE

(Addendum to GFC7000E Manual PN 04584)

TEST deactivates Warning

SAMPLE

 

Messages until New warning(s)

TEST

are activated

 

WHEEL TEMP WARNING

CO = XX.XX

CAL

MSG

CLR SETUP

SAMPLE

RANGE=50.0 PPM

 

 

CO = XX.XX

< TST TST > CAL

MSG

CLR

SETUP

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SAMPLE

WHEEL TEMP WARNING

CO = XX.XX

< TST TST > CAL

MSG

CLR

SETUP

 

 

 

 

 

 

 

Make SURE warning

messages are NOT due to

LEGITIMATE PROBLEMS..

MSG activates Warning

Messages.

<TST TST> keys replaced with

TEST key

Press CLR to clear the

message currently being

Displayed.

If more than one warning is

active the next message will

take its place

Once the last warning has been

cleared, the analyzer returns to

SAMPLE Mode

Figure: Viewing and Clearing Warning Messages

NOTE: A failure of the analyzer’s CPU or Mother Board can result in any or ALL of the following messages.

Table: Warning Messages

Warning

Fault Condition

Possible Causes

Message

 

 

 

 

 

BENCH TEMP

The optical bench

Bench Heater

WARNING

temp is controlled at

Bench Temperature Sensor

 

48 ± 5 °C.

Relay controlling the Bench Heater

 

 

Entire Relay Board

 

 

I2C Bus

BOX TEMP

Box Temp is < 5 °C

Box Temperature typically runs ~7oC warmer than

WARNING

or > 48 °C.

ambient temperature.

 

 

Poor/blocked ventilation to the analyzer.

 

 

Stopped Exhaust-Fan

 

 

Ambient Temperature outside of specified

 

 

range

CANNOT DYN

Dynamic Span

Measured concentration value is too high or

SPAN

operation failed

low.

 

 

Concentration Slope value to high or too low

CANNOT DYN

Dynamic Zero

Measured concentration value is too high.

ZERO

operation failed

Concentration Offset value to high

 

 

 

CONFIG

Configuration and

Failed Disk on Chip

INITIALIZED

Calibration data

User erased data

 

reset to original

 

 

Factory state.

 

 

 

 

DATA

Data Storage in

Failed Disk on Chip

INITIALIZED

iDAS was erased

User cleared data

 

 

 

5

05134 Rev -

Page 5
Image 5
Teledyne GFC7000EU manual Table Warning Messages Fault Condition Possible Causes, Clr

GFC7000EU specifications

The Teledyne GFC7000EU is a state-of-the-art gas flow controller renowned for its precision and reliability in various applications. Designed primarily for the semiconductor industry, this advanced instrument excels in managing and controlling gas flows with unparalleled accuracy. Its compact design allows for seamless integration into existing systems, making it a preferred choice for manufacturers aiming to enhance their production processes.

One of the standout features of the GFC7000EU is its ability to provide high-speed control of gas flow. The controller utilizes advanced algorithms that enable rapid response times, ensuring that gas delivery is not only accurate but also stable under dynamic conditions. This is critical in semiconductor manufacturing, where even slight fluctuations in gas flow can lead to defects in the final product.

The GFC7000EU boasts a wide range of flow capabilities, making it versatile for various gases including toxic, corrosive, and reactive types. Its compatibility with multiple gases allows users to optimize their process without the need for extensive modifications to their setups. Additionally, the controller operates efficiently across a broad temperature range, providing reliable performance in varied environmental conditions.

Another significant aspect of the GFC7000EU is its integration with modern communication technologies. It supports various protocols, including RS-232, RS-485, and Ethernet, allowing for easy connection to automation systems and enhancing overall process monitoring and control. This connectivity is vital for real-time data analysis and remote management, improving operational efficiency.

The GFC7000EU is equipped with advanced diagnostics and self-calibration features that minimize downtime. The device includes built-in troubleshooting capabilities and alerts users to any potential issues, ensuring that maintenance can be performed before problems arise. This proactive approach to equipment management enhances productivity and reduces operational costs.

Constructed with high-quality materials, the GFC7000EU is designed for longevity and durability. Its robust housing protects internal components from environmental factors, making it suitable for use in demanding settings. Overall, the Teledyne GFC7000EU stands out as a premier solution for gas flow management, combining cutting-edge technology with user-friendly features to support the evolving needs of the semiconductor and other industries.