Example:

(A)

What is the output sensitivity of HS-1KS?

 

(B) What is the flow rate for a 2.71 volt output?

Solution:

(A)

Sensitivity

=

rated flow

 

 

 

 

Rated output

 

 

 

=

1000 SCCM

 

 

 

 

5.00 VDC

 

 

 

=

200 SCCM/VDC

 

(B)

Mass Flow

=

output X sensitivity

 

 

 

=

2.71v X 200 SCCM/VDC

 

 

 

=

542 SCCM

5.7Accuracy:

The accuracy of the flow indication is +/- 1% of full scale, with the exception of the U-Series transducer where the accuracy is +/- 3% of full scale.

5.8Repeatability:

The repeatability over a six-month period is +/- ½% of full scale if the flowmeter is operating normally in a

clean, dry system. Under reasonable constant conditions, a repeatability of +/- 1 to+/- 2% of INDICATION can be expected on a day-to-day basis.

5.9Over-range:

The flowmeter has an output signal which is directly proportional to mass flow and linear to +/- of full scale from zero flow to the normal full-scale flow rate.

The flowmeter can be used to measure flow rates higher than the rated maximum flow, but the output signal normally becomes non-linear above 5 volts d-c. A calibration curve for voltage vs. flow can be made for outputs up to 10 volts d-c. At some point above 10 volts d-c the output will no longer increase as the flow increases. The flow rate required to produce this condition is several times the normal full-scale flow rate. Once flow is reduced to the proper level, the flow meter will again indicate mass flow currectly. The flowmeter will not be damaged by excessive flow rates as long as the pressure in the line does not exceed the pressure rating for the transducer.