2.1 EPA DESIGNATION

Model T265 Chemiluminescence Ozone Analyzer Addendum to Model T200 Manual

Specifications, Approvals, &

Certifications

 

2.1 EPA DESIGNATION

Teledyne API’s Model T265 Chemiluminescence Ozone Analyzer received EPA approval for Designation EQOA-0611-199. The designation will be considered valid when operated under the following conditions:

on any full scale range between 0-100 ppb and 0-1000 ppb

with any range mode (Single, Dual, or AutoRange)

at any ambient temperature in the range of 5°C to 40°C

with a TFE filter in the sample air inlet

with a sample flow rate of 500 ± 50 cm3/min (sea level)

with the dilution factor set to 1

with Temp/Press compensation ON

in accordance with the appropriate associated instrument manuals

with or without any of the following options:

Internal or external sample pump

Sample/Cal valve option

Rack mount with or without slides

Analog input option

4-20 mA isolated current loop output

2.2APPROVALS AND CERTIFICATIONS

The Teledyne - API Model T265 analyzer was tested and certified for Safety and Electromagnetic Compatibility (EMC). This section presents the compliance statements for those requirements and directives.

2.2.1 Safety

IEC 61010-1:2001, Safety requirements for electrical equipment for measurement, control, and laboratory use.

CE: 2006/95/EC, Low-Voltage Directive

North American:

cNEMKO (Canada): CAN/CSA-C22.2 No. 61010-1-04

NEMKO-CCL (US): UL No. 61010-1 (2nd Edition)

2.2.2 EMC

EN 61326-1 (IEC 61326-1), Class A Emissions/Industrial Immunity

EN 55011 (CISPR 11), Group 1, Class A Emissions

FCC 47 CFR Part 15B, Class A Emissions

CE: 2004/108/EC, Electromagnetic Compatibility Directive

2.3 WARRANTY

Refer to the Model T200 operation manual regarding the Warranty policy for the Model T265.