Getting StartedModel T265 Chemiluminescence Ozone Analyzer Addendum to Model T200 Manual

 

Step

Action

Comment

 

6.

Press EXIT,

The Model T265 returns to the CAL menu. Now allow

 

 

input Span gas

span gas to flow passed the instrument.

 

 

 

 

 

7.

Wait 30 min

Wait for the O3 reading to stabilize at the span value

 

 

 

(watch O3 STB).

 

8.

Press SPAN

The SPAN button should be displayed once the

 

 

 

concentration approaches the span value. In certain

 

 

 

circumstances at low span concentrations both the

 

 

 

ZERO and SPAN buttons will appear. Do not press

 

 

 

ZERO again when running span gas!

 

 

 

 

 

9.

Press ENTR

Pressing ENTR to span the instrument actually changes

 

 

 

the equations so that the concentration displayed is the

 

 

 

same as the expected span concentration you entered

 

 

 

above, thus spanning the instrument.

 

 

 

 

 

10.

Press EXIT

Pressing EXIT returns the instrument to SAMPLE

 

 

 

mode.

 

 

 

 

IMPORTANT

IMPACT ON READINGS OR DATA

Over time PMTs lose sensitivity, which is reflected by an increase in the O3 slope test function during ongoing calibrations. If the slope falls outside the range of 0.7 to 1.3, a hardware adjustment to the PMT gain is required. The process involves the adjustment of both a high and low gain potentiometer, located on the preamp printed circuit board, while noting the “NORM PMT” test function value, for a particular concentration of Ozone span gas. Typically, the high voltage to the PMT is adjusted, such that, the “NORM PMT” is twice the measured O3 concentration. See PMT Sensor

Hardware Calibration in the Model T200 manual’s Troubleshooting &Service section for detailed procedures.

Step 3 - Review Quality of calibration: Calibration Quality Check Procedure

Step

Action

 

 

1.Scroll the TEST function menu until the O3 SLOPE is displayed.

2.Scroll the TEST function menu until the O3 OFFS is displayed.

Comment

The SLOPE value for O3 should be 1.0 ± 0.3. If the SLOPE value is in the acceptable range the instrument will perform optimally.

The Model T265 will display the offset parameter “O3 OFF” for the O3 equation. This number should be near zero. A value of ±10 indicates calibration in the optimal range.

The Model T265 is now ready to measure sample gas.