Appendix C

Appendix C

Warranty/Repair Questionnaire

M265E and T265

(P/N 074990000 Rev A, DCN# 6167)

CUSTOMER: ____________________________________ PHONE: ___________________________________________

CONTACT NAME: _______________________________ FAX NO. ___________________________________________

SITE ADDRESS: ______________________________________________________________________________________

MODEL SERIAL NO.: ____________________________ FIRMWARE REVISION: _______________________________

1.ARE THERE ANY FAILURE MESSAGES? ____________________________________________________________

_____________________________________________________________________________________________________

_____________________________________________________________________________________________________

PLEASE COMPLETE THE FOLLOWING TABLE: (NOTE: DEPENDING ON OPTIONS INSTALLED, NOT ALL TEST

PARAMETERS SHOWN BELOW WILL BE AVAILABLE IN YOUR INSTRUMENT) *IF OPTION IS INSTALLED

PARAMETER

RECORDED VALUE

ACCEPTABLE VALUE

RANGE

PPB/PPM

100 PPB TO 2000 PPB

O3 STAB

PPB/PPM

< 0.3 PPB (RMS) WITH ZERO

 

 

AIR

SAMPLE FLOW

CM3

500 ± 50

NO FLOW

CM3

5 ± 2

PMT SIGNAL WITH ZERO

MV

-20 TO 150

AIR

 

 

PMT SIGNAL AT SPAN GAS

MV

0-5000MV

CONC

PPB

0-20000 PPB

NORM PMT SIGNAL AT

MV

0-5000MV

SPAN GAS CONC

PPB

0-20000PPB

AZERO

MV

-20 TO 150

HVPS

V

400 – 900

RCELL TEMP

ºC

50 ± 1

 

 

 

BOX TEMP

ºC

AMBIENT + 3 -7ºC

 

 

 

PMT TEMP

ºC

7 ± 2ºC

 

 

 

RCEL PRESS

IN-HG-A

2-10 (constant)

 

 

 

SAMP PRESS

IN-HG-A

AMBIENT - 1 ± 1

NO PRESS

IN PSIG

5 ± 0.2

O3 SLOPE

 

1.0 ± 0.3

O3 OFFSET

 

-20 TO 150

ETEST

PMT MV

2000 ± 1000

 

 

 

OTEST

PMT MV

2000 ± 1000

 

 

 

 

Values are in the Signal I/O

 

REF_4096_MV

MV

4096mv ±2mv and Must be

 

 

Stable

REF_GND

MV

0± 0.5 and Must be Stable

 

 

 

TELEDYNE API CUSTOMER SERVICE

EMAIL: api-customerservice@teledyne.com

PHONE: (858) 657-9800 TOLL FREE: (800) 324-5190FAX: (858) 657-9816

07337B DCN6324

C-1