Appendix C
Warranty/Repair Questionnaire
M265E and T265
(P/N 074990000 Rev A, DCN# 6167)
CUSTOMER: ____________________________________ PHONE: ___________________________________________
CONTACT NAME: _______________________________ FAX NO. ___________________________________________
SITE ADDRESS: ______________________________________________________________________________________
MODEL SERIAL NO.: ____________________________ FIRMWARE REVISION: _______________________________
1.ARE THERE ANY FAILURE MESSAGES? ____________________________________________________________
_____________________________________________________________________________________________________
_____________________________________________________________________________________________________
PLEASE COMPLETE THE FOLLOWING TABLE: (NOTE: DEPENDING ON OPTIONS INSTALLED, NOT ALL TEST
PARAMETERS SHOWN BELOW WILL BE AVAILABLE IN YOUR INSTRUMENT) *IF OPTION IS INSTALLED
PARAMETER | RECORDED VALUE | ACCEPTABLE VALUE |
RANGE | PPB/PPM | 100 PPB TO 2000 PPB |
O3 STAB | PPB/PPM | < 0.3 PPB (RMS) WITH ZERO |
|
| AIR |
SAMPLE FLOW | CM3 | 500 ± 50 |
NO FLOW | CM3 | 5 ± 2 |
PMT SIGNAL WITH ZERO | MV | |
AIR |
|
|
PMT SIGNAL AT SPAN GAS | MV | |
CONC | PPB | |
NORM PMT SIGNAL AT | MV | |
SPAN GAS CONC | PPB | |
AZERO | MV | |
HVPS | V | 400 – 900 |
RCELL TEMP | ºC | 50 ± 1 |
|
|
|
BOX TEMP | ºC | AMBIENT + 3 |
|
|
|
PMT TEMP | ºC | 7 ± 2ºC |
|
|
|
RCEL PRESS | ||
|
|
|
SAMP PRESS | AMBIENT - 1 ± 1 | |
NO PRESS | IN PSIG | 5 ± 0.2 |
O3 SLOPE |
| 1.0 ± 0.3 |
O3 OFFSET |
| |
ETEST | PMT MV | 2000 ± 1000 |
|
|
|
OTEST | PMT MV | 2000 ± 1000 |
|
|
|
| Values are in the Signal I/O |
|
REF_4096_MV | MV | 4096mv ±2mv and Must be |
|
| Stable |
REF_GND | MV | 0± 0.5 and Must be Stable |
|
|
|
TELEDYNE API CUSTOMER SERVICE
EMAIL: api-customerservice@teledyne.com
PHONE: (858)
07337B DCN6324 |