Page 17

12Multimeter Users Manual

Use i only on power supplies and other power sources that have a low output impedance. Do not use i to measure voltage in electronic circuitry unless a 2 ke load will not damage the circuit. See † on page 27.

DISABLING iWITH FUNCTION LOCK

To lock the meter in either the continuity/diode or ohms function, and disable i:

1.Put the slide-switch in the continuity/ohms position. The meter selects the continuity/diode or ohms function. Press [g] to toggle between the continuity/diode and ohms functions.

2.Press [Z] to put the meter in manual range. Z is displayed. The meter is locked in the selected function and i is disabled.

In continuity/diode test, press [Z] to toggle the beeper on and off. In ohms, press [Z] to manually select a range.

To remove the function lock and reenable i, press [Z] for 2 seconds, press [g], or move the slide-switch.

14

Image 17
Contents Multimeter Table of Contents Multimeter Read First Safety Information Symbols Display DisplayOperating Features Operating FeaturesInput Ranges Standby ModeManually Selecting a Range AutorangingMeasuring Voltage Measuring Voltage Testing Continuity and Measuring Resistance Continuity and Resistance Measurements Testing Diodes Testing Diodes Vchek and HOW to USE IT Disabling iWITH Function Lock Measuring Capacitance Measuring Capacitance Recording Minimum and Maximum Readings Using MIN MAX FunctionsDisplaying Minimum and Maximum Reading Recording Minimum and Maximum Reading with Elapsed Time Maximum and Minimum Reading with Elapsed Time Capturing Continuity Intermittents with Continuity Capture Open-to-Short and Short-to-Open Transitions Replacing the Battery MaintenanceTurning Beeper OFF Replacing the Test LeadsSpecifications Service and Parts AccessoriesTerminal and Earth Ground Multimeter Us rs Manual Continuity Capture