Invacare user manual Using the Whirlpool Function, Adjusting the Tub Height 6302GADH Only

Page 28

5 OPERATION

5.6Using the Whirlpool Function

1.If the water level does not cover the whirlpool jets, push the tub fill button to auto-fill the tub.

2.After the water level covers the whirlpool jets and after bathing the resident, add Invacare Defoamer to the bath water. Refer to the Defoamer bottle for directions.

3.Push the whirlpool button to start whirlpool jets.

4.After use, press the whirlpool button again to stop the whirlpool jets.

5.Drain the tub completely.

6.Use the stretcher lift to transfer the resident from the tub.

7.Clean and disinfect the tub. Refer to Cleaning and Disinfecting on page 23.

5.7Adjusting the Tub Height - 6302GADH Only

WARNING

￿DO NOT stand on the tub frame. Keep feet away from tub and frame, otherwise injury may occur.

CAUTION

￿DO NOT adjust the height more than three cycles (up and down) within a ten minute period. Damage to the actuator may result.

DO NOT adjust the tub so it hits the stretcher lift. If necessary, adjust the height of the stretcher lift. Otherwise damage to the tub may occur.

The tub stops automatically when the maximum height adjustment UP or DOWN is reached.

UP Button

DOWN Button

Perform one of the following:

Press and hold the UP button. Release the UP button when the desired height is reached.

Press and hold the DOWN button. Release the DOWN button when the desired height is reached.

Invacare® TheraPure™Supine Whirlpool Tubs

28

Part No. 1167456

Image 28
Contents Page Contents Operation Symbols GeneralLimited Warranty Important Safety Information SafetyGeneral Guidelines Safety Features OverviewWhirlpool Jet Adjustment Flow Control Valve Product Labeling Fixed Height Adjustable Height Typical Product Parameters 6302G 6302GADH6302G 6302GADH Minimum Floor Space and Clearance Requirements Installation Preparing the Work AreaTub Dimensions and Connection Details Connection Location TypeSecuring the Tub Leveling the TubElectrical Installation Plumbing Installation Plumbing Connection DetailsHOT Post Installation Checklist Operation OverviewWater Temperature Control Cleaning and Disinfecting Detail a Detail B Removing Soap Film and Hard Water Deposits Removing Calcium Deposits, Scale and Lime Build-upWater Temperature Setting Using the Shower/Massage Wand Filling the TubUsing the Whirlpool Function Adjusting the Tub Height 6302GADH OnlyInspection, Maintenance and Troubleshooting Safety Inspection ChecklistsInitially Before Each UseSymptoms Faults Solution TroubleshootingHeight Adjustment Model 6302GADH Only Maintenance RecordUSA and Canada Manufacturing Location