4 INSTALLATION
4.6Plumbing Connection Details
3/4 in. Ball Valve | MUST be a | Drain | Wall |
| 38 in. | ||
| minimum of 6 in |
| |
|
| (965 mm) | |
| above the floor | Centerline of Tub | |
|
| ||
|
|
| |
|
| 5 in. |
|
|
| (127 mm) |
|
|
| Tub | Back of Tub |
|
|
| |
|
| Front of Tub | Elbow |
Accommodate Reduced Pressure Zone (RPZ) if required.
2 in. PVC stub |
|
|
| Drain |
|
|
FIGURE 4.1 Plumbing Connection Details
4.7Plumbing Installation
DANGER
Be certain all plumbing work is performed by a qualified plumber and is in compliance with local codes. Check local codes regarding maximum water temperature allowed at hot water outlets for facility type.
WARNINGS
Be certain all plumbing work is performed by a qualified plumber and is in compliance with local codes.
Some state and local plumbing codes require the installation of a “reduced pressure zone” (RPZ) assembly on both the incoming hot and cold supply lines to prevent back flow contamination into the potable water system. Invacare recommends you check local plumbing code requirements to determine if RPZs are required on your tub installation. RPZs are NOT supplied by Invacare and are to be provided by the customer if required.
Invacare® TheraPure™Supine Whirlpool Tubs | 18 | Part No. 1167456 |