Fluke pmn instruction manual     

Page 22

    

WARNING: EXTREME CAUTION MUST ALWAYS BE EXERCISED WHEN

TAKING ANY HIGH VOLTAGE MEASUREMENTS. IT SHOULD BE DONE

ONLY BY QUALIFIED PERSONNEL WHO ARE TRAINED IN THE SAFETY ASPECTS OF WORKING WITH HIGH VOLTAGE.

A sample of the output voltage is available in the REMOTE connector. If it is desired to measure the HV output externally, care must be taken to understand the accuracy of the measurement.

When making a DC measurement, such as when the power supply is holding voltage on a capacitor, any HV probe and DVM combination can be used. The Fluke 80K-40 probe with any 10M input resistance DVM is adequate up to 40kV. Building a simple resistor divider using appropriate HV resistors is also very straightforward. Keep in mind that all HV resistors, including the one in the Fluke probe, exhibit a negative voltage coefficient, changing by up to 4% from zero to max. voltage. Derating the resistors and calibrating at the operating point solves this problem.

The value of the resistor R1 and R2 (Figure 5.3) can be calculated as follows:

V M =

R 2

 

 V O where VO is the High Voltage being measured.

R

R

2

 

1

 

 

Making a pulsed measurement with an oscilloscope requires a compensated HV probe having a wide bandwidth. Simply connecting a DC probe, through the proper resistance, into a scope yields a slow response only adequate for low rep. rate systems. As with DC probes, the pulsed probe resistor voltage coefficient is a problem. In addition, damage to the resistors can occur during pulsing due to high electric field gradients. Also, stray

capacitance to nearby objects can significantly alter the pulse response. For a high-performance, shielded probe to 40KV use a Tektronix P6015 or Ross Engineering VD60-8.3-A-K-LB.

Measurements accurate to better than 0.1% can be achieved using a bias technique. For example, if a 40V signal (40kV divided by 1000) is to be measured accurately, the minus input of the DVM would be biased up 40V. The original signal, with respect to ground, is fed to the plus input of the DVM. The bias can be measured accurately for absolute measurements, or relative measurements read directly as the line or load is varied. In the same manner, an oscilloscope return can be biased for accurate peak measurements during pulsing.

83-493-001 Revision G

5 - 3

Applications

Image 22
Contents    Introduction SpecificationsInstallation ESDMaintenance and Calibration Applications                 #            !     #           Specifications Specifications Specifications Specifications     $          Input AC Power Connection NON-PFC Version      %           Grounding of Input Line Output Ground Connection         Eample Of Interface Connection Revision G Operations Fault   &                      Output Current Measurement                    !            Long Charge Time Operation Revision G Applications