ONLY QUALIFIED PERSONNEL TRAINED IN THE SAFETY ASPECTS OF HIGH VOLTAGE SHOULD PERFORM CALIBRATION.
The calibration steps described in this section require operation of the power supply with the cover removed. Proceed with extreme caution as hazardous voltages are exposed throughout the unit.
Safety glasses must be worn to prevent serious injury in the event of a component failure (e.g., power transistors readily explode during fault conditions).
Because the power supply does not receive proper cooling with the cover removed, it must be cooled by an external fan placed next to the supply to cool the inverter and HV section (min. air flow 100 CFM (2.83 M3/min.)) when operating at full power. Operation at full power with cover removed should be limited to less than five minutes.
No maintenance is required under normal operating conditions. Occasional vacuum or
Calibration of the output voltage is accomplished with trim pot RP4 located on the control board. This is the top PC board of the inverter assembly. RP4 is a 25 turn trimpot. Slowly turn it clockwise to decrease the output voltage for a given VPROGRAM. Factory set for 10V rated voltage for standard version. Refer to Sections 4.1 and 4.2 of product specification for Linearity and Accuracy.
6 - 1 | Maintenance and Calibration |