Time resolution in Sags & Swells and Record modes.

The modes Sags & Swells and Record basically perform similar functions. In both modes Fluke 43B plots a trend of up to two electrical parameters over time. In Sags & Swells mode the instrument plots voltage and current only. In Record mode the analyzer plots a wide variety of parameters, depending on the function that is active when the Record pushbutton is pressed. In both modes the instrument plots the parameters periodically over the so-called plot interval. Refer to the table below for the relationship between Recording Time and Plot Interval. The relationship is based on the fact that there are always 240 plots across the horizontal screen width.

Recording Time

Plot Interval

Recording Time

Plot Interval

 

 

 

 

4 minutes

1 second

8 hours

2 minutes

 

 

 

 

8 minutes

2 seconds

16 hours

4 minutes

 

 

 

 

16 minutes

4 seconds

24 hours

6 minutes

 

 

 

 

30 minutes

8 seconds

48 hours

12 minutes

 

 

 

 

1 hour

15 seconds

4 days

24 minutes

 

 

 

 

2 hours

30 seconds

8 days

48 minutes

 

 

 

 

4 hours

60 seconds

Endless (16 days)

1 to 96 minutes

 

 

 

 

Fluke 43B always measures faster than the plot interval: it always is looking at multiple measurements for each point that is plotted. In fact the analyzer looks at all the measurements it has taken during a plot interval and records a minimum, maximum, and average reading. The difference between Sags & Swells and Record mode is in the measuring rate.

Sags & Swells is optimized for measuring short duration variations of current and voltage. The rms current and voltage of every line cycle are measured. The Fluke 43B then records the min (lowest single cycle), max (highest single cycle), and average measurements at the end of each plot interval.

The Record mode takes measurements roughly 250 milliseconds apart. It also records the min, max, and average during each plot interval.

12

Image 16
Contents Fluke 43B Placing Orders and Getting Assistance Table of Contents Index Carrying Case Contents UnpackingSafety Information Read First Safety Information Read First Fault protection. This is denoted by the symbol Current ProbePowering the Fluke 43B Powering the Fluke 43BOpening Screen Inputs Measurement ConnectionsMain Menu Introducing the Fluke 43B Main Menu Volts / Amps / Hertz PowerHarmonics Transients Sags & SwellsInrush Current Time resolution in Sags & Swells and Record modes Screen with 16 minutes time scale Instrument Setup Setting the DateSetting the Time Adjusting the Contrast Selecting Probes Move to the probe settings for input Setting up the Harmonic- and Power Function Harmonics- and Power Settings Harmonics settings Power settingsSelecting a Language Using FlukeView Software Connecting a ComputerUsing a Printer Connecting a Parallel PrinterSelecting a Printer Type Resetting the Fluke 43B Resetting the Fluke 43BFluke 43B Cleaning and Storage Cleaning the Fluke 43BCleaning the Current Probe Storing the Fluke 43BBatteries Charging the BatteriesExtending Battery Operation Time Replacing the Ni-Cd Battery Pack Replacing the Battery Pack Replaceable Parts Ordering CodesOrdering Code Manuals Item Cont’d Ordering CodeFluke 43B Ordering Code Applications Guide FlukeView User Information Ordering CodeTroubleshooting Safety Characteristics Performance CharacteristicsIntroduction Environmental DataSafety Specifications Maximum input voltage Input 1Function Specifications PowerVolts / Amps / Hertz HarmonicsTransients Sags & SwellsInrush Current Horizontal Input ImpedanceVertical Scope readings DC, a DCTemperature OhmDiode ContinuityMiscellaneous DisplayMemory MechanicalElectrical Specifications Current Probe 80i-500SAccuracy AltitudeHumidity Environmental ConditionsElectromagnetic Compatibility EMC Electromagnetic Immunity Disturbance with test leads TL24 in scope modeDisturbance with test leads TL24 or Current Clamp 80i-500s Declaration of Conformity ManufacturerStatement of Conformity Sample testsLimited Warranty & Limitation of Liability Index Save Key. See Applications Man Ual