Fluke 43B
Users Manual
Safety Specifications
Safety Characteristics
Designed and tested for measurements on 600 V rms Category III, Pollution Degree 2 in accordance with:
•EN/IEC
•ANSI/ISA
•
•
Installation Category III refers to distribution level and fixed installation circuits inside a building.
Maximum input voltage Input 1 and 2 |
|
| |||||||||
Direct on inputs or with test leads TL24 | (see Figure 10) | ||||||||||
0 to 66 kHz | 600 | V rms | |||||||||
> 66 kHz | derating to 5 | V rms | |||||||||
With Shielded | (see Figure 10) | ||||||||||
0 to 400 kHz | 300 | V rms | |||||||||
> 400 kHz | derating to 5 | V rms | |||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Figure 10. Max. Input Voltage v.s. Frequency |
|
Maximum floating voltage |
|
From any terminal to ground |
|
0 to 400 Hz | 600 V rms |
36