Fluke 43B user manual Scope readings, DC, a DC

Page 44

Fluke 43B

Users Manual

Scope readings

The accuracy of all scope readings is valid from 18 °C to 28 °C with relative humidity up to 90 % for a period of one year after calibration. Add 0.1 x (the specified accuracy) for each °C below 18 °C or above 28 °C. More than one waveform period must be visible on the screen.

V DC, A DC

±(0.5 % + 5 counts)

V AC and V AC+DC (True RMS) input 1

 

DC to 60 Hz

±(1 % + 10 counts)

60 Hz to 20 kHz

±(2.5 % + 15 counts)

20 kHz to 1 MHz

±(5 % + 20 counts)

1

MHz to 5 MHz

±(10 % + 25 counts)

5

MHz to 20 MHz

±(30 % + 25 counts)

A AC and A AC+DC (True RMS) input 2

 

DC to 60 Hz

±(1 % + 10 counts)

60 Hz to 15 kHz

±(30 % + 25 counts)

Frequency (Hz), Pulse width, Duty cycle (2.0 % to 98.0 %)

 

1

Hz to 1 MHz

±(0.5 % + 2 counts)

1

MHz to 10 MHz

±(1 % + 2 counts)

10 MHz to 30 MHz

±(2.5 % + 2 counts)

Phase (Input 1 to Input 2)

 

1

Hz to 60 Hz

±2°

60 Hz to 400 Hz

±5°

Peak voltage

 

Peak max, Peak min

± 5 % of full scale

Peak-peak

± 10 % of full scale

Crest

 

Range

1.0 to 10.0

 

 

±(5 % + 1 counts)

40

Image 44
Contents Fluke 43B Placing Orders and Getting Assistance Table of Contents Index Carrying Case Contents UnpackingSafety Information Read First Safety Information Read First Fault protection. This is denoted by the symbol Current ProbePowering the Fluke 43B Powering the Fluke 43BOpening Screen Inputs Measurement ConnectionsMain Menu Introducing the Fluke 43B Main Menu Harmonics PowerVolts / Amps / Hertz Inrush Current Sags & SwellsTransients Time resolution in Sags & Swells and Record modes Screen with 16 minutes time scale Instrument Setup Setting the DateSetting the Time Adjusting the Contrast Selecting Probes Move to the probe settings for input Setting up the Harmonic- and Power Function Harmonics- and Power Settings Harmonics settings Power settingsSelecting a Language Using FlukeView Software Connecting a ComputerUsing a Printer Connecting a Parallel PrinterSelecting a Printer Type Resetting the Fluke 43B Resetting the Fluke 43BFluke 43B Cleaning and Storage Cleaning the Fluke 43BCleaning the Current Probe Storing the Fluke 43BBatteries Charging the BatteriesExtending Battery Operation Time Replacing the Ni-Cd Battery Pack Replacing the Battery Pack Ordering Code Ordering CodesReplaceable Parts Manuals Item Cont’d Ordering CodeFluke 43B Ordering Code Applications Guide FlukeView User Information Ordering CodeTroubleshooting Safety Characteristics Performance CharacteristicsIntroduction Environmental DataSafety Specifications Maximum input voltage Input 1Function Specifications PowerVolts / Amps / Hertz HarmonicsInrush Current Sags & SwellsTransients Vertical Input ImpedanceHorizontal Scope readings DC, a DCTemperature OhmDiode ContinuityMiscellaneous DisplayMemory MechanicalElectrical Specifications Current Probe 80i-500SAccuracy AltitudeElectromagnetic Compatibility EMC Environmental ConditionsHumidity Disturbance with test leads TL24 or Current Clamp 80i-500s Disturbance with test leads TL24 in scope modeElectromagnetic Immunity Declaration of Conformity ManufacturerStatement of Conformity Sample testsLimited Warranty & Limitation of Liability Index Save Key. See Applications Man Ual