Fluke 43B user manual Function Specifications, Volts / Amps / Hertz, Power, Harmonics

Page 41

Specifications 3

Function Specifications

Function Specifications

For all specifications, probe specifications must be added.

Electrical functions

Specifications are valid for signals with a fundamental between 40 and 70 Hz.

Minimum input voltage

4 V peak-peak

Minimum input current

10 A peak-peak (1 mV/A)

Input bandwidth

DC to 15 kHz (unless specified otherwise)

Volts / Amps / Hertz

 

 

 

Readings

V rms (AC+DC), A rms (AC+DC), Hz

Voltage ranges (auto)

5.000 V to 500.0 V, 1250 V

 

 

±(1 % + 10 counts)

Current ranges (auto)

50.00 A to 500.0 kA, 1250 kA

 

 

±(1 % + 10 counts)

Frequency range

10.0 Hz to 15.0 kHz

40.0 to 70.0 Hz

±(0.5 % + 2 counts)

CF Crest Factor range

1.0 to 10.0 ±(5 % + 1 counts)

Power

 

 

 

(1 and 3 Phase, 3 Conductor balanced loads)

 

Readings

Watt, VA, VAR, PF, DPF, Hz

Watt, VA, VAR ranges (auto)

250 W to 250 MW, 625 MW, 1.56 GW

when selected: total (%r):

 

±(2 % + 6 counts)

when selected: fundamental (%f):

±(4 % + 4 counts)

DPF

0.00 to 1.00

0.00 to 0.25

not specified

0.25 to 0.90

 

± 0.04

0.90 to 1.00

 

± 0.03

PF

0.00 to 1.00, ± 0.04

Frequency range

10.0 Hz to 15.0 kHz

40.0 to 70.0 Hz

±(0.5 % + 2 counts)

Harmonics

 

 

 

Number of harmonics

DC..21, DC..33, DC..51

Readings / Cursor readings

 

 

 

V rms / I rms

fund. ±(3 % +

2 counts)

31st ±(5 % + 3 counts), 51st ±(15 % +

5 counts)

Watt

fund. ±(5 % + 10 counts)

31st ±(10 % + 10 counts), 51st ±(30 % +

5 counts)

Frequency of fundamental

 

± 0.25 Hz

37

Image 41
Contents Fluke 43B Placing Orders and Getting Assistance Table of Contents Index Unpacking Carrying Case ContentsSafety Information Read First Safety Information Read First Current Probe Fault protection. This is denoted by the symbolPowering the Fluke 43B Powering the Fluke 43BOpening Screen Measurement Connections InputsMain Menu Introducing the Fluke 43B Main Menu Harmonics PowerVolts / Amps / Hertz Inrush Current Sags & SwellsTransients Time resolution in Sags & Swells and Record modes Screen with 16 minutes time scale Setting the Date Instrument SetupSetting the Time Adjusting the Contrast Selecting Probes Move to the probe settings for input Setting up the Harmonic- and Power Function Power settings Harmonics- and Power Settings Harmonics settingsSelecting a Language Connecting a Computer Using FlukeView SoftwareConnecting a Parallel Printer Using a PrinterSelecting a Printer Type Resetting the Fluke 43B Resetting the Fluke 43BFluke 43B Cleaning the Fluke 43B Cleaning and StorageCleaning the Current Probe Storing the Fluke 43BCharging the Batteries BatteriesExtending Battery Operation Time Replacing the Ni-Cd Battery Pack Replacing the Battery Pack Ordering Code Ordering CodesReplaceable Parts Item Cont’d Ordering Code ManualsFluke 43B Ordering Code Applications Guide FlukeView User Information Ordering CodeTroubleshooting Performance Characteristics Safety CharacteristicsIntroduction Environmental DataMaximum input voltage Input 1 Safety SpecificationsPower Function SpecificationsVolts / Amps / Hertz HarmonicsInrush Current Sags & SwellsTransients Vertical Input ImpedanceHorizontal DC, a DC Scope readingsOhm TemperatureDiode ContinuityDisplay MiscellaneousMemory MechanicalCurrent Probe 80i-500S Electrical SpecificationsAccuracy AltitudeElectromagnetic Compatibility EMC Environmental ConditionsHumidity Disturbance with test leads TL24 or Current Clamp 80i-500s Disturbance with test leads TL24 in scope modeElectromagnetic Immunity Manufacturer Declaration of ConformityStatement of Conformity Sample testsLimited Warranty & Limitation of Liability Index Save Key. See Applications Man Ual