Fluke 875 manual Introduction, General Description, Specifications, Containment Monitor, Detector

Page 5

Introduction

General Description 1

Section 1

Introduction

1.1 General Description

Containment Monitor 875

NOTE

The 875 Containment Monitor has been qualified for use in Nuclear Safety related applications. Qualification Test Reports, 950.301, 950.308A, and 950.310A define the parameters that have been verified by test and are available for purchase.

Containment Monitor 875 functions as an accident monitor for reactor containments, refer to Figure 1-1. It consists of the 877-1 Ion Chamber Detector that is located within the reactor containment, and the 876A-1 Readout Module that is contained in a rack in the control room. The readout module contains a power supply that provides the necessary power both for itself and the detector. The readout module is connected to the detector by two cables, a coaxial high voltage cable and a coaxial signal cable. Specifications for the detector and readout module are contained below.

1.2 Specifications

Detector 877-1

Radiation Detected

Photons above 60 keV

Range

1 R/h to 1E7 R/h

Energy Response

± 20% from 80 keV to 2 MeV

Chamber Construction

Stacked parallel plate, 316 stainless steel housing, 3

 

terminal, guarded ion chamber

Saturation

>80% up to full scale on instrument

Collection Voltage

450 to 550 VDC

Current Output

7E-11 to 7E-4 Amp/R/Hr (nominal)

Chamber Capacitance

435 picofarads (nominal)

Voltage Withstand and Leakage

5E10-5 amperes @ 500 VDC between either ground and

 

connector pins

Capacitance

Approximately 430 pf

1-1

Image 5
Contents High Range Containment Monitor VictoreenFluke Biomedical Table of Contents Calibration and Test Procedures Introduction SpecificationsGeneral Description Containment MonitorReadout Module 876A-1 Temperature Operating 40 to 120F 4 to 49CRack Chassis Equipment OverviewHigh-Range Containment Monitor Detector Cables/Connectors/Panel Pull Box AssemblyFlexible Tubing 878-12-30TAB Optional EquipmentHigh-Range Containment Monitor Typical Energy Response Curve Detector Typical Linearity Detector Receiving Inspection High-Range Detector, Recommended InstallationStorage Procedures, Warnings, and CautionsBlank Installation InstallationReadout Module Pull BoxCable Sealing Specifications for CableDetector Test After Installation Cable and Wiring InstallationSpecifications for Cable Detector Cable Inside ContainmentAll Alarms on P2 Normal Operation Shelf State Ancillary WiringPower Connections P1 Computer & RecorderOperation OperationVictoreen Readout Module 876A-1 Function DescriptionFunctional Description Negative Power Supply Operation of High Voltage Supply During ECS Test High Voltage Power Supply Normal Operation non-ECS TestFAIL/SAFE Comparison Circuit on Power Supply Board Alert Alarm Circuit High Alarm CircuitPen Clamp Fail/Safe CircuitRecorders and Computer Buffers Meter at extreme lower end of scale Panel Meter Action During ECS TestMonitoring Circuits of The ECS Board Action of Green Safe/Reset Light and Alarm During ECS TestLow Voltage Ramp Generator Block Diagram of ECS Test Muting Circuits of ECS Board Safe/Reset Sub-circuitsBlank Calibration Maintenance, Calibration, and TroubleshootingMaintenance Test Procedure TP876A-1-108 Troubleshooting Mother Board P/N 876-1-78, Schematic 876-1-3A Signal Input CircuitPreamplifierIMeter Board P/N 876A-1-86, Schematic 876A-1-3E Rear PanelTest Point Normal Channel Test TP501 TP502 +6.3Normal Low Signal Level Channel Test High Alarm CircuitMuting Stages of the Alarm Circuits ESC Board P/N 876A-1-92, Schematic 876A-1-3D Wave Shapes on the Electronic Check Source Board Starting With Terminal H on Relay Driver Board Starting With Pins M and P on the Power Supply SchematicOverall Fail Circuitry Associated with the ECS Test Operation of the Latch Circuit Outputs of U206 Pulse Generator on the ECS BoardCircuit assumed initially in Pass Condition Fail Condition Latch CircuitPin Voltage State Comments NORCalibration and Test Procedures Blank Cable and Pull Box Procedures Cable and Pull Box ProceduresVictoreen Document Number Description Applicable Drawings and Bill of MaterialsApplicable Drawings Detector Cable Out of Containment Applicable Bill of Materials Fluke Biomedical