Fluke 875 manual Maintenance, Calibration, and Troubleshooting

Page 31

Maintenance, Calibration, and Troubleshooting

Maintenance 5

Section 5

Maintenance, Calibration, and Troubleshooting

5.1 Maintenance

The monitor is designed to operate for long periods in the containment without attention. The following replacement schedule should keep the monitor in trouble-free operation.

Every five years - Replace the 4200 microfarad electrolytic capacitor, P/N 92-3005-A (C1 01), and the RFI line filter, P/N 92-9015A (FL101), located on the 876-1-78 mother board of the 876A-1-108 or 876A-100 Readout Module.

Whenever detector cables are removed - Replace the nickel seals, P/N 877-1-60-1, on the detector connectors as described in the CABLE-877 procedure in Appendix A.

5.2 Calibration

The monitoring systems require calibration before placing them into service. In addition, they should be recalibrated at regular intervals during routine service. The length of time between calibration intervals should be determined by operations personnel. For further calibration information, refer to the applicable calibration procedure provided in Appendix A.

The high-range containment area monitor underwent a complete electronic and isotopic calibration prior to leaving the Fluke Biomedical plant. The same electronic calibration procedure is supplied in this manual. Prior to primary isotopic calibration, the detector's hermetically must be verified. Primary isotopic calibration requires a highly radioactive source (greater than 400 curies) with National Bureau of Standards (NBS) traceability. As this is beyond the capability of most facilities to perform, the following method of verifying detector calibration is used:

Detectors shall either be returned to Fluke Biomedical at a five (5) year interval from the date of delivery or the owner must establish a procedure to determine that the average A/R/h output current does not deviate from original factory calibration by more than ± 10%.

To encompass Nureg-0737 guidelines, on-site in-situ calibration checks can be performed with the 878- 10 High-Range Field Calibrator that is capable of producing a 10 R/h indication on the channel under test.

Electronic alignment of the 876A Readout may be performed using Functional Test Procedure, TP876A- 1-108 included in Appendix A, and the standard test equipment as listed in Table 5-1. Personnel must be fully knowledgeable in the operation of the readout prior to attempting to perform the test.

NOTE

Disconnect Detector 877-1 from Readout Module 876A-1 by removing cables from J1 and J2 before starting electronic calibration.

5-1

Image 31
Contents High Range Containment Monitor VictoreenFluke Biomedical Table of Contents Calibration and Test Procedures Containment Monitor SpecificationsIntroduction General DescriptionReadout Module 876A-1 Temperature Operating 40 to 120F 4 to 49CHigh-Range Containment Monitor Detector Equipment OverviewRack Chassis Optional Equipment Pull Box AssemblyCables/Connectors/Panel Flexible Tubing 878-12-30TABHigh-Range Containment Monitor Typical Energy Response Curve Detector Typical Linearity Detector Receiving Inspection High-Range Detector, Recommended InstallationStorage Procedures, Warnings, and CautionsBlank Pull Box InstallationInstallation Readout ModuleCable Sealing Specifications for CableDetector Cable Inside Containment Cable and Wiring InstallationDetector Test After Installation Specifications for CableComputer & Recorder Ancillary WiringAll Alarms on P2 Normal Operation Shelf State Power Connections P1Operation OperationVictoreen Functional Description Function DescriptionReadout Module 876A-1 Negative Power Supply FAIL/SAFE Comparison Circuit on Power Supply Board High Voltage Power Supply Normal Operation non-ECS TestOperation of High Voltage Supply During ECS Test Alert Alarm Circuit High Alarm CircuitRecorders and Computer Buffers Fail/Safe CircuitPen Clamp Meter at extreme lower end of scale Panel Meter Action During ECS TestLow Voltage Ramp Generator Action of Green Safe/Reset Light and Alarm During ECS TestMonitoring Circuits of The ECS Board Block Diagram of ECS Test Muting Circuits of ECS Board Safe/Reset Sub-circuitsBlank Maintenance Maintenance, Calibration, and TroubleshootingCalibration Test Procedure TP876A-1-108 Troubleshooting Rear Panel Signal Input CircuitMother Board P/N 876-1-78, Schematic 876-1-3A PreamplifierIMeter Board P/N 876A-1-86, Schematic 876A-1-3ETest Point Normal Channel Test TP501 TP502 +6.3Muting Stages of the Alarm Circuits High Alarm CircuitNormal Low Signal Level Channel Test ESC Board P/N 876A-1-92, Schematic 876A-1-3D Wave Shapes on the Electronic Check Source Board Overall Fail Circuitry Associated with the ECS Test Starting With Pins M and P on the Power Supply SchematicStarting With Terminal H on Relay Driver Board Operation of the Latch Circuit Outputs of U206 Pulse Generator on the ECS BoardCircuit assumed initially in Pass Condition Fail Condition Latch CircuitPin Voltage State Comments NORCalibration and Test Procedures Blank Cable and Pull Box Procedures Cable and Pull Box ProceduresVictoreen Applicable Drawings Applicable Drawings and Bill of MaterialsDocument Number Description Detector Cable Out of Containment Applicable Bill of Materials Fluke Biomedical