Fluke 875 manual Outputs of U206 Pulse Generator on the ECS Board, Operation of the Latch Circuit

Page 40

Victoreen 875

Operators Manual

2.An auxiliary pulse from pin 3 of U208 to the automatic pulse generating circuits resets them so that an automatically generated pulse will be present 17.1 minutes later if the ECS Test is not conducted manually during that time. The purpose of this pulse is to insure continued operation of automatic testing.

5.15 Outputs of U206 Pulse Generator on the ECS Board

We have noted that U206 can be triggered by pressing the ECS button. Its main output is a power pulse to the ramp generator. However, it also initiates a pulse from U203 (pin 10) to pin 5 of U207A, the muting multivibrator. The muting multivibrator also has two outputs:

1.One output to U203 that ultimately causes a muting pulse six seconds wide at terminal 9. This is coupled to the power supply board on pin M. It also returns to the circuit so that jitters will be eliminated on manual ECS.

2.A second output to U207 is a one-shot multivibrator that provides an enabling pulse to the latch circuit (which also contains a flip-flop). The latch circuit consists of four NOR gate sections of U202 (pins 1 through 6 and 8 through 13), and also inverter U208 (pins 9 and 10). The latch circuit output is taken from terminal 8, and will be either 6.2 V or 14 V. The operation is explained below.

5.16 Operation of the Latch Circuit

The latch circuit has two inputs and one output:

Input 1 - Enters on pin 6 (U202) and pin 10 (U208), which are connected. This is the monitoring input. Its value is the result of the ECS Test. It enables the circuit to change state only after the conditions brought about by the ECS Test have stabilized.

Input 2 - Enters pin 1 (U202) and pin 5 of U202 which are connected. The output is taken from pin 10 and pin 2 of U202, which are connected.

Output - The output may be either 0 V, which is a pass indication, or +14 VDC, which is a fail indication. The two NOR gates (pins 8, 9, 10, and pins 11, 12, 13) are effectively a flip-flop circuit which will change its state only when a certain combination of the two inputs is present. The presence of CR201 and a pull-up circuit on the power supply printed circuit board permits pin 10 to be 0 while terminal 8 is 6.2 V.

Input 1 has a normal resting voltage of +15 V. However, during the ECS Test it takes on a voltage determined by the steady detector current that exists during the ECS Test. This current generates a voltage of about 2.57 V (0.33 scale) in the panel meter, at which time pin 6 of U202 is at approximately 0 V. If, at this instant, the enabling voltage (input 2) arrives, it will cause the flip-flop to switch states.

Input 2 is initiated by pressing the ECS button, but its actual occurrence in time is delayed so that it occurs during the flat portion of the current cycle generated by the ECS ramp.

Assume that the green light is ON at the beginning of the ECS Test. Actually this is probable because:

If the green light were off, troubleshooting would not begin with the ECS Test. With this assumption we can work backwards in the truth table (Table 5-1) from the output (which must be low) to the possible inputs.

In Figure 5-2, the initial output is assumed low (circuit condition did not show fail prior to ECS Test).

When the enable pulse arrives in the two cases of the signal pulse [pass (L) or fail (H)] the following conditions exist:

U202, pins 12, 10 L (Assumed as initial condition)

5-10

Image 40
Contents Victoreen High Range Containment MonitorFluke Biomedical Table of Contents Calibration and Test Procedures Specifications IntroductionGeneral Description Containment MonitorTemperature Operating 40 to 120F 4 to 49C Readout Module 876A-1High-Range Containment Monitor Detector Equipment OverviewRack Chassis Pull Box Assembly Cables/Connectors/PanelFlexible Tubing 878-12-30TAB Optional EquipmentHigh-Range Containment Monitor Typical Energy Response Curve Detector Typical Linearity Detector High-Range Detector, Recommended Installation Receiving InspectionProcedures, Warnings, and Cautions StorageBlank Installation InstallationReadout Module Pull BoxSpecifications for Cable Cable SealingCable and Wiring Installation Detector Test After InstallationSpecifications for Cable Detector Cable Inside ContainmentAncillary Wiring All Alarms on P2 Normal Operation Shelf StatePower Connections P1 Computer & RecorderOperation OperationVictoreen Functional Description Function DescriptionReadout Module 876A-1 Negative Power Supply FAIL/SAFE Comparison Circuit on Power Supply Board High Voltage Power Supply Normal Operation non-ECS TestOperation of High Voltage Supply During ECS Test High Alarm Circuit Alert Alarm CircuitRecorders and Computer Buffers Fail/Safe CircuitPen Clamp Panel Meter Action During ECS Test Meter at extreme lower end of scaleLow Voltage Ramp Generator Action of Green Safe/Reset Light and Alarm During ECS TestMonitoring Circuits of The ECS Board Block Diagram of ECS Test Safe/Reset Sub-circuits Muting Circuits of ECS BoardBlank Maintenance Maintenance, Calibration, and TroubleshootingCalibration Test Procedure TP876A-1-108 Troubleshooting Signal Input Circuit Mother Board P/N 876-1-78, Schematic 876-1-3APreamplifierIMeter Board P/N 876A-1-86, Schematic 876A-1-3E Rear PanelTP501 TP502 +6.3 Test Point Normal Channel TestMuting Stages of the Alarm Circuits High Alarm CircuitNormal Low Signal Level Channel Test ESC Board P/N 876A-1-92, Schematic 876A-1-3D Wave Shapes on the Electronic Check Source Board Overall Fail Circuitry Associated with the ECS Test Starting With Pins M and P on the Power Supply SchematicStarting With Terminal H on Relay Driver Board Outputs of U206 Pulse Generator on the ECS Board Operation of the Latch CircuitLatch Circuit Circuit assumed initially in Pass Condition Fail ConditionNOR Pin Voltage State CommentsCalibration and Test Procedures Blank Cable and Pull Box Procedures Cable and Pull Box ProceduresVictoreen Applicable Drawings Applicable Drawings and Bill of MaterialsDocument Number Description Detector Cable Out of Containment Applicable Bill of Materials Fluke Biomedical