Agilent Technologies 3138A-00101 and above Location and Cooling, Bench Operation, Rack Mounting

Page 14

Figure 2-2. Power Cord Installation

Location and Cooling

Bench Operation

Chapter 1 gives the dimensions of your mainframe. The mainframe has plastic feet that are shaped to ensure self-alignment when stacked with other Agilent System II cabinets. The feet may be removed for rack mounting. Your mainframe must be installed in a location that allows sufficient space at the sides and rear of the mainframe for adequate air circulation. Do not block the vents at the sides and rear of the mainframe.

Rack Mounting

The mainframe can be mounted in a standard 19-inch rack panel or cabinet. Refer to Chapter 1 for mainframe dimensions. A fully loaded mainframe can weigh over 36 kg (80 lbs). Install the mainframe into the rack before installing the modules into the mainframe.

Rack mounting kits are available as option numbers 908 and 909 (with handles). Installation instructions are included with each rack mounting kit.

Note

Instrument support rails are also required for rack mounting. These are normally supplied with the

 

cabinet and are not included with the rack mounting kits.

 

 

Temperature Performance

The mainframe can operate within the temperature range of 0° to 55° C. Fans in each module cool the modules by drawing in air through the sides of the mainframe and exhausting it out the back. Using Agilent rack mount or slide kits will not impede the flow of air.

Installation 15

Image 14
Contents Installation Guide Certification Safety Summary Safety Summary Declaration of Conformity Printing HistoryTable Of Contents Figures TablesIntroduction Safety ConsiderationsGeneral Information Instrument IdentificationOptions and Parts DescriptionSpecifications and Supplemental Characteristics Specifications and Supplemental CharacteristicsINH/FLT Characteristics Maximum Ratings Input Power InstallationInstallation Items SuppliedPower Cord Installation FusesBench Operation Temperature PerformanceLocation and Cooling Rack MountingRear Panel Switches Frame Mode SwitchSwitch Settings Installation Module Installation Module RemovalMainframe Turn-On Checkout Module Connections ConnectionsRear Panel Keyboard Connection MPS Mainframe/Power Supply ConnectionsController Connections Digital ConnectionsController Connections Digital Connector Trigger ConnectionsExamples of the Remote Inhibit Function Connections Index Agilent Sales and Support Offices United States Latin AmericaManual Updates