Agilent Technologies 3138A-00101 and above manual Specifications and Supplemental Characteristics

Page 10

Specifications and Supplemental Characteristics

Table 1-1 lists the specifications and supplemental characteristics of the MPS mainframe. Module specifications are included in the appropriate module User’s Guide. Specifications are warranted in the temperature range of 0 to 55 degrees C. Supplemental characteristics are not warranted, but are descriptions of minimum, maximum, or average performance that has been determined by design or type testing.

Table 1-1. Specifications and Supplemental Characteristics

PERFORMANCE SPECIFICATIONS

AC Input Ratings

Standard:87- 132 Vac

174- 250 Vac

Frequency tolerance:

47 - 63 Hz

SUPPLEMENTAL CHARACTERISTICS

Maximum AC Line Current Ratings (with 8 power supply modules operating at maximum output)

100 Vac:

29 A rms

120 Vac:

25 A rms

200 Vac:

16 A rms

220/230 Vac:

16 A rms

240 Vac:

15 A rms

Maximum power input:

3200 VA; 1800 W maximum; 1600 W typical

 

 

 

GPIB Interface Capabilities

Languages:

Standard Commands for Programmable Instruments (see module

 

Programming Guide)

IEEE 488.2 Capabilities:

SH1, AH1, T6, L4, SR1, RL1, PPO, DC1, DT1, El, and CO

 

 

Serial Link Capabilities (Multiple mainframes sharing one GPIB address)

Number of mainframes: Number of modules: Serial-linked power supplies:

Chain length:

TrigIn/TrigOut Characteristics

Maximum Ratings:

Trig Out (HC TTL output):

Trig In (10k pull-up):

2

16

total number of modules and serial-linked power supplies (Agilent 667xA/665xA/664xA/603xA) cannot exceed 16

30 m (100 ft) maximum

±16.5 V between center conductor and chassis

VOL @ 4 ma = 1.34 V maximum

VOL @ 2 ma = 0.8 V maximum

VOH 4 ma = 2.66 V minimum

VOH 2 ma = 3.18 V minimum

IIL @ 4 V = 700 μA

IIH @ 5.25 V = 200 μA

VIL = 0.8 V maximum

VIH = 2.0 V minimum

General Information 11

Image 10
Contents Installation Guide Certification Safety Summary Safety Summary Declaration of Conformity Printing HistoryTable Of Contents Figures TablesIntroduction Safety ConsiderationsGeneral Information Instrument IdentificationOptions and Parts DescriptionSpecifications and Supplemental Characteristics Specifications and Supplemental CharacteristicsINH/FLT Characteristics Maximum Ratings Input Power InstallationInstallation Items SuppliedPower Cord Installation FusesBench Operation Temperature PerformanceLocation and Cooling Rack MountingRear Panel Switches Frame Mode SwitchSwitch Settings Installation Module Installation Module RemovalMainframe Turn-On Checkout Rear Panel ConnectionsModule Connections Keyboard Connection MPS Mainframe/Power Supply ConnectionsController Connections Digital ConnectionsController Connections Digital Connector Trigger ConnectionsExamples of the Remote Inhibit Function Connections Index Agilent Sales and Support Offices United States Latin AmericaManual Updates