Rear Panel Switches

The six-position rear panel switch is shown in the following figure.

Figure 2-3. Rear Panel Switches

Address Switches

Address switches A0 through A4 set the primary GPIB address. The address is set in binary, A0 being the least significant bit and A4 being the most significant bit. Figure 2-3 shows the factory-set address of "5" (binary 00101). Any address from decimal 0 to decimal 30 is a valid GPIB address. Refer to the following diagram for switch settings.

Address

0

1

2

3

4

5

6

7

8

9

A4

0

0

0

0

0

0

0

0

0

0

Switch Positions

A3

A2

A1

0

0

0

0

0

0

0

0

1

0

0

1

0

1

0

0

1

0

0

1

1

0

1

1

1

0

0

1

0

0

 

 

 

A0

0

1

0

1

0

1

0

1

0

1

Address

10

11

12

13

14

15

16

17

18

19

A4

0

0

0

0

0

0

1

1

1

1

Switch Positions

A3

A2

A1

1

0

1

1

0

1

1

1

0

1

1

0

1

1

1

1

1

1

0

0

0

0

0

0

0

0

1

0

0

1

 

 

 

A0

0

1

0

1

0

1

0

1

0

1

Address

20

21

22

23

24

25

26

27

28

29

30

A4

1

1

1

1

1

1

1

1

1

1

1

Switch Positions

A3

A2

A1

0

1

0

0

1

0

0

1

1

0

1

1

1

0

0

1

0

0

1

0

1

1

0

1

1

1

0

1

1

0

1

1

1

A0

0

1

0

1

0

1

0

1

0

1

0

Frame Mode Switch

The frame mode switch determines the IEEE 488 secondary addresses (also referred to as subaddresses) of the modules that are installed in the mainframe. The frame mode switch must be set to MAIN when the mainframe is connected directly to the GPIB. With the frame mode switch set to MAIN, modules are assigned secondary addresses of 0 to 7, which corresponds to the slot locations indicated on the mainframe.

The frame mode switch must be set to AUX (auxiliary) when the mainframe is serially-linked to another mainframe that is directly connected to the GPIB. This also applies when the mainframe is serially linked to an Agilent 603xA/664xA/665xA/ 667xA power supply that is directly connected to the GPIB.

In mainframes that have the frame mode switch set to AUX, modules are assigned secondary addresses of 8 to 15. Secondary address 8 corresponds to slot location 0 on the mainframe and secondary address 15 corresponds to slot location 7 on the mainframe as shown in the following diagram.

16Installation

Image 15
Contents Installation Guide Certification Safety Summary Safety Summary Printing History Declaration of ConformityTable Of Contents Tables FiguresInstrument Identification Safety ConsiderationsGeneral Information IntroductionDescription Options and PartsSpecifications and Supplemental Characteristics Specifications and Supplemental CharacteristicsINH/FLT Characteristics Maximum Ratings Items Supplied InstallationInstallation Input PowerFuses Power Cord InstallationRack Mounting Temperature PerformanceLocation and Cooling Bench OperationFrame Mode Switch Rear Panel SwitchesSwitch Settings Installation Module Removal Module InstallationMainframe Turn-On Checkout Connections Rear PanelModule Connections Digital Connections MPS Mainframe/Power Supply ConnectionsController Connections Keyboard ConnectionController Connections Trigger Connections Digital ConnectorExamples of the Remote Inhibit Function Connections Index United States Latin America Agilent Sales and Support OfficesManual Updates