Xantrex Technology RS232-XPD manual Communications, Baud Rate, Flow Switch Setting Condition

Page 22

Installation and Configuration

Communications

Communications

Baud Rate Serial transmission sends and receives data in bit streams at fixed bit rates. Both the Selection computer and the interface must have the same bit rate setting for proper

communication. Use switches B1, B2, and B3 on the rear panel S1 switch to select the power supply's rate of transmission in bits per second. Table 2.2, “Switch S1 Settings for Baud Rate”summarizes the switch settings with the reference markings on the rear panel subplate. Refer to Figure 2.2, on page 17 for the location of the S1 switch on the RS-232 interface subplate.

Table 2.2 Switch S1 Settings for Baud Rate

Baud Rate

B3

B2

B1

9600

1

1

1

 

 

 

 

4800

0

1

1

 

 

 

 

2400

1

0

1

 

 

 

 

1200

0

0

1

 

 

 

 

600

1

1

0

 

 

 

 

300

0

1

0

 

 

 

 

150

1

0

0

 

 

 

 

75

0

0

0

 

 

 

0 = OPEN/OFF

1 = CLOSED/ON

 

Note:When resetting switches B3, B2, and B1 during a working session, turn the power supply OFF and then ON again to ensure that the new settings are put into effect.

Flow Control Flow control signals regulate data flow for proper communication. Use the FLOW Selection switch on the rear panel S1 switch to select flow control or to disable flow control.

With flow control enabled, you can use either software or hardware protocols to control flow rates. Table 2.3 gives the FLOW switch settings with regards to the markings on the subplate rear panel. For a diagram of the rear panel subplate, see Figure 2.2, on page 17.

Table 2.3 Flow Control Section

FLOW Switch Setting

Condition

1

Disable Flow Control

 

 

0

Enable Flow Control

 

 

20

Operating Manual for RS-232 for XPD Series Power Supply

Image 22
Contents RS232-XPD Page Internal RS-232 Interface for XPD 500 Watt Series How do you get service? Phone 604 422 Toll Free North America 800 667 Fax 604 421What does this warranty not cover? Disclaimer Product Information About Your Power Supply Model Number Serial Number Purchased From Purchase DateSafety SupplyMain Sections Manual RevisionsAbout This Manual Contents Contents Features and Specifications DescriptionFeatures and Specifications Features and FunctionsFeatures Readback FunctionsSpecifications Models 18-30 33-16 60-9 120-4.5Features and Specifications Introduction Initial InspectionInstallation and Configuration Initial InspectionRS-232 Interface Subplate Located on Power Supply Rear Panel RS-232 Interface PCB Basic Setup Procedure Step # Description Action ReferenceCommunications Flow Switch Setting ConditionBaud Rate XON Switch Setting Condition Jumper Connections RS-232 Connector PinPower Supply DCE Host Computer DTE Pin connector Jumper J10 Connections RS-232 Connector PinRemote Switch Position Power on resultsStartup Condition Default Settings Model Example IsetExample RS-232 Connector J4 Pinouts default configuration RS-232 ConnectionPower Supply Settings Parameter SettingUser Options and Settings 12 OVP Control Mode SelectionHigh OFF LOW User Signals User Signals J21 ConnectorJ21 User Signal Connector Circuit Block Diagram J21 User Cable with Ferrite Block Command Syntax RS-232 OperationOperation FormatParameter Description Form Floating Number Definition ExampleOperation Command Summary Command DescriptionAUXA? Cmode Command Reference Auxa 1 onResponse DLY seconds Lockout. See also Remote/Local Operation on Snsp Remote/Local Operation on Status, and Fault Registers on Section Accumulated Status, Status, and Fault Registers Condition Mnemonic Bit Position Bit Weight 4096Error Codes Error # Error Identification ExplanationTroubleshooting LEDsCalibration Voltage Program Calibration Procedure Voltage Mode CalibrationSetup VoltageVoltage Readback Calibration Procedure Current Calibration Setup Current Mode CalibrationCurrent Program Calibration Procedure Current Readback Calibration Procedure Over Voltage Protection OVP Calibration Page Xantrex Technology Inc