Section 4. Calibration

Introduction

WARNING

Exercise caution when using and servicing power supplies. High energy levels can be stored at the output voltage terminals on all power supplies in normal operation. In addition, potentially lethal voltages exist in the power circuit and the output connector of power supplies which are rated at 40V and over. Filter capacitors store potentially dangerous energy for some time after power is removed.

You can calibrate the RS-232 interface by adjusting the signal levels on the interface card so that they correspond to the expected signal levels on the power supply's main assembly. You may need to recalibrate the interface if you replace parts either on the interface board or on the main power supply board, or if the unit falls out of specification due to component aging drifts.

You can calibrate the RS-232 Interface for:

Voltage program

Voltage readback

Current program

Current readback

Overvoltage protection

The following equipment will be required to accurately calibrate your unit:

Digital Voltmeter, 5 1/2 digit, 0.1% accuracy or better, with test leads

Current sensing shunt resistor, rated for 150% of maximum output current, 0.25% accuracy or better

Connection wires rated for the unit's maximum output current and voltage

An RS-232 equipped computer to send the calibration commands to the unit

Calibrate the unit according to the following procedures, referring to “Command Reference” for more information about the calibration commands used.

The calibration procedures in this section are designed to be performed at an ambient temperature of 25°C ± 5°C.

Release 1.2

51

Image 53
Contents RS232-XPD Page Internal RS-232 Interface for XPD 500 Watt Series Phone 604 422 Toll Free North America 800 667 Fax 604 421 How do you get service?What does this warranty not cover? Disclaimer Product Model Number Serial Number Purchased From Purchase Date Information About Your Power SupplySupply SafetyManual Revisions Main SectionsAbout This Manual Contents Contents Description Features and SpecificationsFeatures Features and FunctionsFeatures and Specifications Readback FunctionsModels 18-30 33-16 60-9 120-4.5 SpecificationsFeatures and Specifications Initial Inspection IntroductionInitial Inspection Installation and ConfigurationRS-232 Interface Subplate Located on Power Supply Rear Panel RS-232 Interface PCB Step # Description Action Reference Basic Setup ProcedureBaud Rate Flow Switch Setting ConditionCommunications Jumper Connections RS-232 Connector Pin XON Switch Setting ConditionJumper J10 Connections RS-232 Connector Pin Power Supply DCE Host Computer DTE Pin connectorStartup Switch Position Power on resultsRemote Iset Condition Default Settings Model ExampleExample RS-232 Connection RS-232 Connector J4 Pinouts default configurationParameter Setting Power Supply Settings12 OVP Control Mode Selection User Options and SettingsHigh OFF LOW User Signals J21 Connector User SignalsJ21 User Signal Connector Circuit Block Diagram J21 User Cable with Ferrite Block RS-232 Operation Command SyntaxFormat OperationFloating Number Definition Example Parameter Description FormOperation Command Description Command SummaryAUXA? Cmode Auxa 1 on Command ReferenceResponse DLY seconds Lockout. See also Remote/Local Operation on Snsp Remote/Local Operation on Status, and Fault Registers on Section Accumulated Status, Status, and Fault Registers 4096 Condition Mnemonic Bit Position Bit WeightError # Error Identification Explanation Error CodesLEDs TroubleshootingCalibration Setup Voltage Mode CalibrationVoltage Program Calibration Procedure VoltageVoltage Readback Calibration Procedure Current Program Calibration Procedure Current Mode CalibrationCurrent Calibration Setup Current Readback Calibration Procedure Over Voltage Protection OVP Calibration Page Xantrex Technology Inc