Xantrex Technology RS232-XPD manual Current Mode Calibration, Current Calibration Setup

Page 56

Calibration

Current Mode Calibration

Current Mode Calibration

Current

Calibration

Setup

1.Disconnect the load from the power supply to be calibrated.

2.Connect a shunt across the supply's output terminals.

3.Connect a voltmeter across the shunt.

Power Supply

Power Supply

Positive Output

Negative Output

Current Sensing

Resistor

Current

Program

Calibration

Procedure

+V -

VOLTMETER

Figure 4.2 Current Calibration Setup

1.Connect the shunt and voltmeter to the power supply as shown in Figure 4.2.

2.Activate calibration mode by sending command CMODE ON or CMODE 1 to the power supply.

3.Send command ILO; VLO to the power supply. Measure and record the output shown on the external voltmeter.

4.Send command IHI; VHI to the supply Measure and record the output voltage shown on the external voltmeter.

5.Calculate ILO and IHI from the voltages read from the external voltmeter and the shunt resistance. I=V/R.

6.Send the command IDATA <ilo>,<ihi> to the power supply. <ilo> and <ihi> are the current values obtained from sending the ILO and IHI commands to the power supply. When the power supply is calibrated, the low to high current program calibration values are stored as constants.

7.Program the supply at various levels using the ISET command to confirm that the calibration was successful and that linearity is observed. Refer to the current program accuracy specification in Section 1.

8.Turn off calibration mode by sending the command CMODE OFF or CMODE 0 to the power supply.

54

Operating Manual for RS-232 for XPD Series Power Supply

Image 56
Contents RS232-XPD Page Internal RS-232 Interface for XPD 500 Watt Series How do you get service? Phone 604 422 Toll Free North America 800 667 Fax 604 421What does this warranty not cover? Disclaimer Product Information About Your Power Supply Model Number Serial Number Purchased From Purchase DateSafety SupplyMain Sections Manual RevisionsAbout This Manual Contents Contents Features and Specifications DescriptionFeatures and Functions FeaturesFeatures and Specifications Readback FunctionsSpecifications Models 18-30 33-16 60-9 120-4.5Features and Specifications Introduction Initial InspectionInstallation and Configuration Initial InspectionRS-232 Interface Subplate Located on Power Supply Rear Panel RS-232 Interface PCB Basic Setup Procedure Step # Description Action ReferenceBaud Rate Flow Switch Setting ConditionCommunications XON Switch Setting Condition Jumper Connections RS-232 Connector PinPower Supply DCE Host Computer DTE Pin connector Jumper J10 Connections RS-232 Connector PinStartup Switch Position Power on resultsRemote Condition Default Settings Model Example IsetExample RS-232 Connector J4 Pinouts default configuration RS-232 ConnectionPower Supply Settings Parameter SettingUser Options and Settings 12 OVP Control Mode SelectionHigh OFF LOW User Signals User Signals J21 ConnectorJ21 User Signal Connector Circuit Block Diagram J21 User Cable with Ferrite Block Command Syntax RS-232 OperationOperation FormatParameter Description Form Floating Number Definition ExampleOperation Command Summary Command DescriptionAUXA? Cmode Command Reference Auxa 1 onResponse DLY seconds Lockout. See also Remote/Local Operation on Snsp Remote/Local Operation on Status, and Fault Registers on Section Accumulated Status, Status, and Fault Registers Condition Mnemonic Bit Position Bit Weight 4096Error Codes Error # Error Identification ExplanationTroubleshooting LEDsCalibration Voltage Mode Calibration SetupVoltage Program Calibration Procedure VoltageVoltage Readback Calibration Procedure Current Program Calibration Procedure Current Mode CalibrationCurrent Calibration Setup Current Readback Calibration Procedure Over Voltage Protection OVP Calibration Page Xantrex Technology Inc