Fisher FB300-2 manual Getting Started, Using the Power Supply

Page 6

Getting Started

Select a location that allows for 3" clearance behind the power supply, and comfortable reach of the front panel controls and cell connections. Do not block the vented area of the case - on the front bottom of the unit, or the fan area at the rear. Connect the power supply to a 3-prong grounded AC outlet, using the AC cord provided with the unit only. Connect the electrophoresis apparatus to the power supply, making sure to match the red positive lead to the red positive jack, and the black negative lead to the black negative jack. Similarly, connect one or two more electrophoresis apparatus to the power supply if you plan to run more than one in parallel. Power the unit on using the AC power switch located at the rear of the unit, next to the AC line cord entry.

black negative jacks

 

3 digit LED display

 

up/down arrows

select key (Set)

constant key (K)

run key

red positive jacks

 

 

 

gray ON/OFF key

 

 

Figure 1. Front Panel Controls

Using the Power Supply

Press the gray key on the front center of the unit to enable the control logic. The 3-digit display will illuminate and show the setpoint value of the last saved run setup. The FB300 preserves the run settings each time you start a run. One set of conditions is saved for each mode, constant voltage or constant current. The setpoint value, limit parameter, time duration, and power-fail restart setting is saved. This allows convenient setup for repetitive runs.

To change the control mode, press the constant key. This key toggles control from constant voltage to constant current. Each time, the FB300 will display the last saved setpoint value for that particular mode.

After selecting a control mode, either constant voltage or current, and setting the limit and timed run parameters if desired (see Constant Voltage/Current Operations) press the Run key to energize the power supply output. The power supply output will ramp up to the appropriate setpoint, while not allowing the limit parameter to be exceeded. If the limit parameter should be exceeded, then the power supply will crossover control modes, making the limit parameter the new control setpoint. The front panel constant mode LED indicator will change to indicate that a crossover has occurred.

FB300-2 MANUAL

Page 5

7608001 Rev A 8/02

Image 6
Contents FB300-2 Electrophoresis Power Supply Operator’s ManualFisher Biotech FB300-2 Electrophoresis Power Supply Table of ContentsSafety Considerations Unpacking the Power Supply IntroductionDC output Power SpecificationsAC input Power EnvironmentalUsing the Power Supply Getting StartedConstant Voltage Operations Constant Current OperationsTimed Operations Automatic Power-fail Restart CleaningTroubleshooting and Error Indications Replacing a fuseFB300-2 Warranty Statement ComplianceAccessories Replacement PartsTable des matières Considérations de sécurité Déballage de lalimentation électrique Environnement SpécificationsCourant dentrée CA Courant de sortie CCMise en route Utilisation de lalimentation électriqueOpérations à tension constante Opérations à intensité constanteOpérations à durée déterminée Redémarrage automatique en cas de coupure de courantNettoyage Remplacement dun fusible Dépannage et indications derreurGarantie du FB300 ConformitéPièces de rechange Fusible, T1,5A,250VFisher Biotech FB300-2 Elektrophorese-Netzteil InhaltsverzeichnisSicherheitstechnische Hinweise Einführung Auspacken des NetzteilsUmfeldbedingungen Technische DatenNetzeingangsleistung GleichstromausgangsleistungEinstieg Verwenden des Netzteils Betrieb bei DauerspannungBetrieb bei Dauerstrom Zeitgesteuerter BetriebAutomatischer Neustart bei Stromausfall Reinigung Fehlersuche und Fehlermeldung Auswechseln einer SicherungFB300 Garantieerklärung KompatibilitätErsatzteile Fisher Scientific Park Lane Pittsburgh, PA NetzkabelSicherung, T1,5A Contenido Consideraciones de seguridad Introducción Desembalaje de la fuente de alimentaciónMedioambientales EspecificacionesPotencia de entrada de CA Potencia de salida de CCPara empezar Uso de la fuente de alimentaciónOperaciones a voltaje constante Operaciones a corriente constanteOperaciones temporizadas Reinicio automático tras un fallo en la alimentaciónLimpieza Diagnóstico de problemas e indicaciones de error Sustitución de un fusibleDeclaración de Garantía del FB300 ConformidadPiezas de repuesto Fusible, T1.5A,250VFB300-2 Electrophoresis Power Supply Part No Rev a