Fisher FB300-2 manual Constant Voltage Operations, Constant Current Operations, Timed Operations

Page 7

To stop a run in progress, press the Run key or the gray ON/OFF key. The display will change to "OFF" signifying that the output is no longer energized. Pressing run again resumes the run from the previous point (i.e. for a timed run). Pressing any other key will return to setup mode in either constant voltage or constant current mode as appropriate.

During a run, the display will automatically cycle between voltage, current, and time displays. Voltage and current is displayed as actually measured by the FB300. When displaying the controlled setpoint, i.e. voltage for constant voltage control, the arrow keys can be used to modify the setpoint. The limit parameter cannot be changed during a run. The time parameter displays elapsed time for an untimed run, and time left for a timed run.

Constant Voltage Operations

Press the constant key (labeled K) until the LED indicating "Volts" is illuminated. Each time the constant key is pressed, the FB300 will display the last saved setpoint value for that particular mode. Use the UP and DOWN arrow keys to change the setpoint to the desired value (10-300 volts). The FB300 will not allow setting a voltage setpoint outside of the allowed range. Press and hold either arrow key to quickly change the setpoint.

The limit parameter, in this case current, is normally set to the power supply maximum value of 400 mA. To set a lower limit value, press the Set key until the LED indicating "mAmps" is illuminated. Use the UP and DOWN arrow keys to change the limit to the desired value (4-400 mA). The FB300 will not allow setting a current limit outside of the allowed range.

In addition, the FB300 will never exceed the maximum power output specification of 75W. When starting a run, the power supply calculates a maximum limit parameter, above which the power supply maximum power output rating would be exceeded. If this value is lower than the limit setpoint entered, the FB300 will use this lower limit parameter.

Constant Current Operations

Press the constant key (labeled K) until the LED indicating "Amps" is illuminated. Each time the constant key is pressed, the FB300 will display the last saved setpoint value for that particular mode. Use the UP and DOWN arrow keys to change the setpoint to the desired value (4-400 mA). The FB300 will not allow setting a current setpoint outside of the allowed range. Press and hold either arrow key to quickly change the setpoint.

The limit parameter, in this case voltage, is normally set to the power supply maximum value of 300 volts. To set a lower limit value, press the Set key until the LED indicating "Volts" is illuminated. Use the UP and DOWN arrow keys to change the limit to the desired value (10-300 volts). The FB300 will not allow setting a voltage limit outside of the allowed range.

In addition, the FB300 will never exceed the maximum power output specification of 75W. When starting a run, the power supply calculates a maximum limit parameter, above which the power supply maximum power output rating would be exceeded. If this value is lower than the limit setpoint entered, the FB300 will use this lower limit parameter.

Timed Operations

It is possible to enter an amount of time for the power supply to provide power, after which it will automatically shut off. Use this feature for timed runs.

Press the Set key until the LED indicating "Time" is illuminated. Use the UP and DOWN arrow keys to change the time duration to the desired value (0-999 minutes). A time duration of zero effectively disables timed run mode.

When running a timed run, the time parameter displays the time remaining in the run. When running a non-timed run, the time parameter displays the elapsed time during the run

FB300-2 MANUAL

Page 6

7608001 Rev A 8/02

Image 7
Contents Operator’s Manual FB300-2 Electrophoresis Power SupplyTable of Contents Fisher Biotech FB300-2 Electrophoresis Power SupplySafety Considerations Introduction Unpacking the Power SupplyEnvironmental SpecificationsAC input Power DC output PowerGetting Started Using the Power SupplyConstant Current Operations Constant Voltage OperationsTimed Operations Cleaning Automatic Power-fail RestartReplacing a fuse Troubleshooting and Error IndicationsCompliance FB300-2 Warranty StatementReplacement Parts AccessoriesTable des matières Considérations de sécurité Déballage de lalimentation électrique Courant de sortie CC SpécificationsCourant dentrée CA EnvironnementUtilisation de lalimentation électrique Mise en routeOpérations à intensité constante Opérations à tension constanteRedémarrage automatique en cas de coupure de courant Opérations à durée déterminéeNettoyage Dépannage et indications derreur Remplacement dun fusibleConformité Garantie du FB300Fusible, T1,5A,250V Pièces de rechangeInhaltsverzeichnis Fisher Biotech FB300-2 Elektrophorese-NetzteilSicherheitstechnische Hinweise Auspacken des Netzteils EinführungGleichstromausgangsleistung Technische DatenNetzeingangsleistung UmfeldbedingungenEinstieg Betrieb bei Dauerspannung Verwenden des NetzteilsZeitgesteuerter Betrieb Betrieb bei DauerstromAutomatischer Neustart bei Stromausfall Reinigung Auswechseln einer Sicherung Fehlersuche und FehlermeldungKompatibilität FB300 GarantieerklärungFisher Scientific Park Lane Pittsburgh, PA Netzkabel ErsatzteileSicherung, T1,5A Contenido Consideraciones de seguridad Desembalaje de la fuente de alimentación IntroducciónPotencia de salida de CC EspecificacionesPotencia de entrada de CA MedioambientalesUso de la fuente de alimentación Para empezarOperaciones a corriente constante Operaciones a voltaje constanteReinicio automático tras un fallo en la alimentación Operaciones temporizadasLimpieza Sustitución de un fusible Diagnóstico de problemas e indicaciones de errorConformidad Declaración de Garantía del FB300Fusible, T1.5A,250V Piezas de repuestoFB300-2 Electrophoresis Power Supply Part No Rev a