Agilent Technologies E2050-90003 manual Troubleshooting Network Configurations

Page 86

Troubleshooting

Troubleshooting Network Configuration

Troubleshooting Network Configurations

If some network configuration values for the E2050 LAN/GPIB Gateway are improperly configured, one or more client systems may not be able to make a connection to the Gateway. The problem may be:

nImproper configuration value(s) for the default subnet gateway and/ or the subnet mask. If this is the case, the Gateway may be receiving requests from the client, but routing the replies to the wrong place.

nImproper or incomplete configuration value(s) for the IP allow list. If the client IP address does not match the ones allowed in the configured IP allow list, the connection will be refused. To verify this:

lCheck for a response from the Gateway with the ping command. If a response occurs, the connection is being refused because of the IP allow list configuration values.

lIf you are using a syslog server, look for access denied messages in the syslog file. See “syslog File Messages” for more information.

Telnet Configuration To correct network configuration values when using the Telnet configuration method, take one of the following actions:

nIf a client can successfully Telnet to the Gateway, use the Telnet configuration commands to fix the values. Then, reboot using the Telnet reboot command.

CAUTION

Any client operations in progress and client connections will be terminated by the reboot command.

nOtherwise, reset the Gateway to the default configuration values by pressing the Config Preset button on the back panel of the LAN/GPIB Gateway at the same time as you power-on the Gateway. Then, see Chapter 3 to reconfigure the Gateway.

86

Chapter 5

Image 86
Contents Page Page Contents 103 Warranty Information Government Restricted RightsCopyright Information Trademark InformationPrinting History Contacting AgilentSafety Considerations Radio and Television InterferencePage Declaration of Conformity Page Introduction Introduction ChapterHow to Use This Guide Guide ContentsRelated Software Documentation For HP-UXE2050 Hardware Description E2050 Hardware DescriptionE2050 LAN/GPIB Gateway Front Panel E2050 LAN/GPIB Gateway Front PanelE2050 LAN/GPIB Gateway Rear Panel E2050 LAN/GPIB Gateway Rear PanelRack Mount Kit Optional AC/DC Adapter and AC Power CordAC/DC Adapter and Standard AC Power Cord Rack Mount Kit Optional ChapterE2050 Software/Firmware Architecture E2050 Software/Firmware ArchitectureApplication Software Supported Typical LAN Connections Typical LAN Connections ChapterClient Server Configuration LANSoftware/Firmware Architecture Overview Software/Firmware Architecture ChapterUsing Application Software Establishing a Network ConnectionInstallation Installation Hardware Requirements Hardware RequirementsSoftware Requirements Software RequirementsFinding the LAN Hardware Address MAC Configure the Network Installing the HardwareInstalling the Hardware Record LAN Hardware AddressInstalling the Hardware Connect the Gateway to the Network Installing the Hardware Installing the Hardware Connect Gpib Cable and Power Cords Connecting the Power Module ChapterInstalling the Hardware Observe the Power-On Sequence Installing the Hardware Chapter Configuration Configuration Configuration Values Descriptions Setting Configuration ValuesSteps to Set Configuration Values Setting Configuration Values0.0 Hpib Allow 156.140.34-48.2 Sec Typically, syslogd obtains its routing information from How Configuration Values are Used When the Config Preset Button is Pressed Configuration Methods Configuration MethodsConfiguration Methods Overview Bootp Using Telnet Configuration Method Windows Route add 192.0.0.192 yoursystemname Enter# Default subnet gateway IP address 120 Ip 156.140.222.201 Enter Configuration Methods Using Telnet Configuration Method HP-UX # Default subnet gateway IP address Telnetcommand value Return Ping 156.140.4.249 Return Packets transmitted, 0 packets received, 100% packet loss Using Bootp Configuration Method HP-UX Be rfc1048 Ping 156.140.4.249 Enter Using Bootp with Tftp Configuration Method HP-UX E2050\ Hostname E2050 Configuration Methods Ping 156.140.4.249 Return Administration Administration Using the Telnet Utility Using the Telnet UtilityAccessing the Telnet Utility Using the Telnet Utility OFF Telnet Commands Exiting the Telnet UtilityAllow IPaddresses Querying the Firmware Revision Querying the Current ConfigurationQuerying the Gateway Querying the GatewayQuerying the Configuration Method Querying Gateway/Client ConnectionsServer ID Igpibantctl Igpibgett1delay Changing the Configuration Method Configuring the GatewaySetting Default Configuration Values Configuring the GatewayConfiguring the Gateway Using the Gateway’s syslog File MessagesMaintaining the syslog File Terminating Client Connections Configuring the Gateway Chapter Troubleshooting Troubleshooting Gateway Failure Messages Sicl Error Codes and MessagesGateway Failure Messages Syslog File Messages IPaddressVerifying Network Connections Troubleshooting Network ConfigurationTroubleshooting Network Configuration Troubleshooting Network Configurations Bootp or Bootp with Tftp Configuration Setting Default Configuration Troubleshooting Gateway ConfigurationVerifying Gateway Configuration Troubleshooting Gateway ConfigurationVerifying the Configuration Method LAN Conn LED Description OperationTroubleshooting Gateway Configuration Changing the Configuration Method Troubleshooting Telnet Configuration Troubleshooting Bootp ConfigurationTroubleshooting Bootp with Tftp Configuration Troubleshooting Client Connections Troubleshooting Client ConnectionsClient Connection Problems Packets transmitted, 0 packets received, 100% packet loss Client Run-time Errors LAN/GPIB Gateway followTroubleshooting Client Connections Troubleshooting Client Connections Specifications Specifications 100 SpecificationsSpecifications 101 Power RequirementsGpib Characteristics Environmental102 Specifications Glossary 103Glossary 104 GlossaryGlossary 105 106 Glossary Index Index 107108 Index 109 110