Agilent Technologies Performance Verification, 1a. Agilent E1364A Test Fixture Schematic

Page 21

Performance Verification

The procedures in this section are used to test the switch module’s electrical performance using the specifications in Appendix A of the E1364A User’s Manual as the performance standards. These tests are suitable for incoming inspection, troubleshooting, and preventive maintenance.

Test Fixture A test fixture is required for the following tests. It is recommended that you order an extra terminal module (aka "terminal block") to use as a test fixture, so that you do not have to re-wire the terminal module each time these tests are performed. The terminal module part number is E1364-80001.

Figures 2-1(a) and 2-1(b) show how the test fixture should be wired.

Perform the following steps to wire the test fixture:

Short all NO (Normally Open) lines together.

Short all NC (Normally Closed) lines together.

Short all C (Common) lines together.

Figure 2-1(a). Agilent E1364A Test Fixture Schematic

Chapter 2

Verification Tests 21

Image 21
Contents Serial Numbers Agilent E1364A Channel Form C Switch ModulePage Contents Appendix B. Backdating Information Limitation Of Warranty WarrantyCertification Exclusive RemediesDocumentation History Safety SymbolsManufacturer’s Address Manufacturer’s NameDeclares, that the product Product Name Model NumberPage Page Page Introduction Agilent E1364A Form C Switch ModuleGeneral Information Chapter Safety ConsiderationsChapter General Information Initial Incoming Inspection Guidelines Inspection/ShippingPackaging/Shipping Guidelines Switch Description Temperature HumidityEnvironment Requirements Recommended Model Use Recommended Test EquipmentGeneral Information Chapter Verification Tests Operation Verification Functional VerificationPerformance Verification 1a. Agilent E1364A Test Fixture Schematic1b. Agilent E1364A Test Fixture Test 2-1 Closed-Channel Resistance Test Normally Closed Resistance Test Next 530 540 END Test 2-2 DC Isolation Test Normally Closed to Chassis 1. Hardware Connections DC Isolation Test NC Channels to Chassis Example DC Isolation Test Verification Tests Chapter Test Accuracy Ratio TAR Performance Test RecordPerformance Test Record for the E1364A Switch Module Page 1 Performance Test Record for the E1364A Switch Module Page 2 Maximum Uncert Test Measured Description MinimumTest Measured Description Minimum Reading Maximum Uncert Performance Test Record for the E1364A Switch Module Page 4Verification Tests Chapter Replaceable Parts Lists Part Qty Part DescriptionNumber Replaceable Parts Chapter E1364A Terminal Block Replaceable Parts Replaceable Parts Chapter Service Identifying the Problem Troubleshooting TechniquesE1364A Tests/Checks Test/Check Reference Designator Testing the AssemblyESD Precautions Repair and Maintenance GuidelinesFunctional Verification Test Verification Tests C ProgramsExample Closed Channel Resistance Test Performance Verification TestsClose instrument session Example DC Isolation Test Iclose dm Example C Programs Appendix a Reference Part Qty Part Description Designator Number Replaceable Parts ListQty Part Description Part NumberAgilent E1364A Reference Designators Figure B-1. E1364A Component Assembly Parts Figure B-2. E1364A Terminal Module Replaceable Parts Page Page Copyright Agilent Technologies, Inc